MALFUNCTION MESSAGES
5
Malfunction messages
Malfunctions
The following malfunction messages are displayed on the control unit, the message is
displayed with further information as full text in the SOPAS ET operating program.
NOTICE:
Malfunctions can cause damage to the device.
Message
Significance
Possible cause
Measure
CB4
●
Fuse CB4 of the heated
extraction hose has triggered
or has been manually
deactivated.
●
Current too high or manually
deactivated.
●
Check power circuit for short
circuit.
●
Check power consumption.
●
Switch fuse back on.
CB5
●
Fuse CB5 of the purge air
heater has triggered or has
been manually deactivated.
●
Current too high or manually
deactivated.
●
Check power circuit for short
circuit.
●
Check current consumption.
●
Switch fuse back on.
Water purging configura-
tion
●
Water purging not activated.
●
Hardware activation not com-
pleted or software key not
entered.
●
Enter software key or contact
SICK Service.
THeiz3 control
●
Heater 3 control deviation.
●
Temperature or other control
value not reached.
●
Check fuse.
●
Check nominal temperature.
●
Check settings.
THeiz4 control
●
Heater 4 control deviation.
●
Temperature or other control
value not reached.
●
Check fuse.
●
Check nominal temperature.
●
Check settings.
Tmax Heiz3
●
Maximum temperature set
reached.
●
Temperature reached.
●
Limit value set too low.
●
Flow rate too low.
●
Temperature reached.
●
Limit set too low.
●
Flow rate too low.
Tmax Heiz4
●
Maximum temperature set
reached.
●
Temperature reached.
●
Limit value set too low.
●
Flow rate too low.
●
Adapt limit value.