6.4.3
Diagnostics Comparison Report Check
To compare the current diagnostic values (current fingerprint) of the meter with the
reference diagnostic values (reference fingerprint, e.g. stored at commissioning), the
Diagnostics Comparison Report provides graphs and color coded information on
diagnostics.
Figure 86
Diagnostics Comparison Log table
To print a Diagnostics Comparison Report, complete the following steps:
Download the Diagnostics Comparison Log from your FLOWSIC600 according to
.
Go to the Diagnostics Comparison data tab (
Click the button "Preview/Print".
In the "Preview / Print options", enter your name and select the flow direction for which
you want to view the report.
Click "Preview / Print".
In the print preview, either print the report or export it to a PDF file.