Maintenance
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
177
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
6.4.1.2
Acknowledging logbook entries on the meter
To acknowledge logbook entries on the meter, proceed as follows:
Download and save the logbook entries from the meter according to
.
Select the logbook in which entries are to be acknowledged or select "All logbooks" to
acknowledge entries in all logbooks at once.
Mark the entries to be acknowledged.
Click the "Acknowledge selection" button to acknowledge the selected entries only or
click the "Acknowledge all" button in order to acknowledge all entries in the selected
logbook(s).
6.4.1.3
Clearing logbooks on the meter
If the logbooks are configured with the storage behavior "rolling", it is not necessary to
clear the logbooks on the meter. When the logbook is full, new entries will overwrite the
oldest entries.
If a logbook is configured with the storage behavior "blocking" (e.g. with custody configura-
tion), a full Custody logbook [1] will activate the meter status "Data invalid". In this case it
is recommended to clear the logbooks.
To clear the logbooks on the meter, proceed as follows:
Choose the User Access Level "Service" (
).
Download and save the logbook entries from the meter according to
.
Select the logbook to be cleared or select "All logbooks" to clear all.
Switch the meter to Configuration Mode (choose "File / Configuration Mode" from the
menu).
Click the "Clear meter logbook" button and confirm the warning with "OK".
Switch the meter to Operation Mode.
If the Parameter write lock was unlocked prior to clearing the meter logbook, follow all
necessary procedures to bring the meter to back to its original state.
NOTICE:
The following preconditions must be met to clear logbooks on the meter:
●
The Parameter write lock must be in the "UNLOCKED" position (
●
The user must be in the User Access Level "Service" (password see Service
Manual).
●
The meter must be in Configuration Mode.
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...