5.2.2
Connecting the FLOWSIC600 via RS485/USB converter
If the PC/laptop does not offer a RS232 serial interface, a USB interface can be used. An
appropriate USB converter is necessary to transform the signal for the RS485 device
interface. The USB converter available through SICK contains a CD-ROM with a software
driver which must be installed before an online connection between the FLOWSIC600 and
the MEPAFLOW600 CBM software can be established.
Figure 55
Wiring example of "MEPA interface set RS485/USB" (Converter, Cable, Terminal plug, CD-ROM with
software driver), non-intrinsically safe (see also Fig. 104)
*Possible terminals for the RS485 connection are:
●
33 (+) and 34 (-)
●
81 (+) and 82 (-)
If necessary, the RS485 ports can be assigned to a specific bus address (Reg.
#5020 "DeviceBusaddress"). The service port always has the bus address "1".
Terminals*
Connection cable
Safe zone
RS485/USB
Table 17
Interface sets for the connection of the FLOWSIC600 to a network
Description
Part Number
"MEPA interface set RS485 / RS232" intrinsically safe for DIN rail
mounting
2033410
"MEPA interface set RS485/USB" (Converter, cable, terminal plug, CD-
ROM with software driver), non-intrinsically safe
6030669
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...