FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH
Assembly and installation
3.3.2.1
Cable specification (connection of a measuring point)
Recommended cable types for on-site connection of MCU and junction box
1
UNITRONIC LiYCY (TP) 2 x 2 x 0.5 mm
²
1 pair of wire for RS485, 1 pair of wire for transducer power supply;
Not suitable for underground laying (protected installation required if necessary)
2
UNITRONIC Li2YCYv (TP) 2 x 2 x 0.5 mm
²
1 pair of wire for RS485, 1 pair of wire for transducer power supply; can be used as an
alternative to Pos. 1; Suitable for underground laying
3
UNITRONIC Bus FD P Combi 3 x 2 x 0.25 + 3 x 1.0 mm
2
1 pair of wire for RS485, 1 pair of wire with cross-section 1.0 mm
2
for transducer power
supply, for cable lengths up to 1000 m with 2 measuring points connected
4
Special cable type ASS 4 x 2 x 0.5 mm
²
1 pair of wire for RS485, 1 to 3 pairs of wires for transducer power supply;
Silicone, halogen-free, highly heat and cold resistant, sheath color red (similar to RAL
3000)
Accessories: Braided cable sleeving PA-S 4, black, to provide mechanical protection or
to cover the sheathing color if necessary.
Connecting
Data transfer
Power supply
FLSE200
Cable type
Sender/receiver unit and
junction box (SICK scope of
delivery)
FLSE200 Master connection:
2 pairs of wires (twisted pair), operating
capacitance wire/wire < 110 pF/m,
wire cross-section
≥
0.25 mm
2
1 pair of wires with
wire cross-section
1 mm
2
UNITRONIC FD P BUS Combi
3 x 2 x 0.25 mm
2
+ 3 x 1 mm
2
FLSE200 Slave connection:
1 pair of wires (twisted pair), operating
capacitance wire/wire < 110 pF/m,
wire cross-section
≥
0.25 mm
2
MCU and junction box
(on-site wiring)
Junction box - junction box
(on-site wiring)
1 pair of wires (twisted pair), operating
capacitance wire/wire < 110 pF/m,
wire cross-section
≥
0.25 mm
2
1 pair of wires with
wire cross-section
≥
0.5 mm
2
(AWG20)
E.g. UNITRONIC
Li2YCYv (TP) 2 x 2 x 0.5 mm
2
or equivalent
Manufacturer of UNITRONIC cables: LAPP-Kabel
Manufacturer of special cable: metrofunk KABEL-UNION GmbH
Cables from other manufacturers can be used when they have the same
transmission characteristics.
We recommend coordinating the wiring of a control unit with several
measuring points (up to 8) with the manufacturer (
For standard wiring (bus wiring), the factory-set termination must be
disabled in the system components not at the end of the line.
NOTICE:
The manufacturer does not assume any warranty for the proper
functioning of the device when cables that do not comply with the
specifications are used.
The cables are to be laid continuously without changing to other cables
or lines and shielded throughout.
Not suitable are telephone cables such as type A-2YF(L)2Y... .
Содержание FLOWSIC200
Страница 1: ...O PERATING INSTRU CTIO NS FLOWSIC200 Flow Velocity Monitor Installation Operation Maintenance...
Страница 30: ...30 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Product description...
Страница 100: ...100 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Commissioning and parameterization...
Страница 104: ...104 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Maintenance...
Страница 121: ...Specification FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH 121...