COMMISSIONING
5
5.3.2
Temperature transmitter
1
Go to menu item “Configuration/Run/Temperature”.
2
Set the “Temperature pressure input type” to “Analog input”.
3
Go to menu item “IO/Diagnostics/Analog inputs”.
“Analog input 2 input value” must be between 4 mA and 20 mA.
If this is not the case, the function of the transmitter must be checked and whether
there are any error values defined by the device.
Adjusting the scale in the flow computer to the working range of the transmitter:
1
Go to menu item “IO/Diagnostics/Analog inputs“.
2
Set the value for “Analog input 2 full scale” to the maximum value of the measuring
range of the temperature transmitter.
3
Set the value for “Analog input 2 zero scale” to the minimum value of the measuring
range of the temperature transmitter.
5.4
Device configuration and connection check
5.4.1
FLOWSIC600-XT
Checking the communication status
1
Go to menu item “Communication/Flowsic600-XT Flow meter/Flowsic600-XT
Communication”
2
Check the “Communication status”.
When the “Communication Status” is set to “OK”, the correct device ID is already set in
the flow computer.
Fig. 29: Menu item “
Communication/Flowsic600-XT Flow Meter/FLOWSIC600-XT Communication”