“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
107
11.4
Standby
Function
The “Standby” condition can be used if you wish to interrupt measurement selectively or if
the sample gas should temporarily not be supplied to the EuroFID. In comparison to
shutting down the device, the “Standby” status subjects the measuring system to less
strain, as all components remain at operating temperature and it is quicker to resume
measurement mode.
“Standby” mode entails the following:
– The sample gas supply is switched over internally to instrument air, i.e. the measuring
system is continuously purged with zero gas.
– The fuel gas supply to the FID is internally interrupted, the FID flame extinguished.
– The “Measured value valid” status output is deactivated.
– The “OK” LED does not light up.
– The ejector pump and the heater of the sensor block remain in operation.
Procedure
1
Open the menu: Main menu
Maintenance
Standby
2
Press
or .
➝
The selection box is highlighted.
3
P r e s s
until the required number is displayed
(
< Y E S >
or
< N O >
).
4
Press
or .
➝
< E n d
>
is highlighted.
5
Press .
➝
The EuroFID switches to the selected condition. (After switchover from
Y E S
to
N O
, the FID flame is automatically ignited.)
➝
The “Maintenance” menu is displayed (
).
S t a n d b y
S t a n d b y
< N O >
C a l i b . G a s S a m p l e G a s
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Selection box
Current measured value
Current measuring range (measuring range end value)
If you open the measurement display (
), the “Standby” condition
is automatically deactivated.
< E n d >
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...