Start-up and Parameter Settings
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
75
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Figure 63
Assigning the MCU to the sender/receiver unit
4.4.2
Factory settings
*: Only when an optional analog module is present (standard on DUSTHUNTER T100 and T200)
The steps required to modify these settings are described in the following Sections.
Prerequisite: The device data are located in the "Project Tree" window, the Level 1
password has been entered and "Maintenance" mode set.
Parameter
Value
Function control
Every 8 h; output of control values (every 90 s) on
standard analog output
Analog output (AO)
[mA]
Live zero (LZ)
4
Upper measuring range value
(MBE)
20
Current during maintenance
0,5
Current by malfunction
21 (optional 1)
Response time
60 s for all measured variables
Measured variable
Output on AO
Value at LZ
Value at MBE
Transmission [%]
100
0
Opacity [%]
1
0
100
Extinction
3 *
0
2
Dust concentration [mg/m³]
2 *
0
200
Regression coefficients (only for dust concentration)
0.00 / 1.00 / 0.00