99
8020216/19YN/V3-0/2020-12| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K
Subject to change without notice
MAINTENANCE
12
12.5.2
Checking the contamination value
▸
Start the function check. To do this, select device file “MCU” and move it to window
“Project Tree”. Switch to directory “Adjustment / Function Check - Manual”. Click “Start
Manual Function Check”.
Fig. 55: SOPAS ET menu: MCU/Adjustment/Function Check - Manual
▸
Select the “DH SP100” device file in the “Project Tree” window, select the “Diagnosis /
Check values” directory and check the contamination value.
Fig. 56: SOPAS ET menu: DH SP100/Diagnosis/Check values
▸
When the measured values for contamination, zero point and span are within the
allowed ranges, save them to the device by clicking the “Refresh” button (“Check values”
field); if not, repeat cleaning and check the contamination value again by triggering a
renewed function check.
▸
Reassemble the sender/receiver unit. Remove the cover from the flange with tube (blind
flange). Fit the sender/receiver unit on the duct.
▸
Resume measuring operation (
see “Resuming measuring operation”, page 94
The function check can also be started via the LC display on the MCU(DH Ex) (
).
●
The contamination value can also be displayed on the LC display of the
MCU(DH Ex) (initiate a function check and switch to the “SP100/Diagnosis”
menu,
●
The device is probably defective when the contamination value does not sink
below the warning value (30%) despite several cleaning processes
→
contact
SICK Service.