O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-2K
Subject to change without notice
13
TROUBLESHOOTING
13.5
Control unit malfunctions
13.5.1
Malfunctions
Table 25: Control unit malfunctions
13.5.2
Warning and malfunction messages
Malfunctions listed below can probably be cleared onsite.
Table 26: Control unit malfunctions that can be cleared
Malfunction
Possible cause
Action
No display on the LCD
●
No supply voltage
●
Connection line to LC display not connected
or damaged
●
Defective fuse
▸
Check voltage supply.
▸
Check connection line.
▸
Exchange fuse.
▸
Contact SICK Service.
Message
Significance
Possible cause
Action
AO configura-
tion
The number of available and
configured analog outputs is not
identical.
●
No parameters set for AO
●
Connection error
●
Module failure
▸
Check configuration (
log outputs parameters”, page 74
▸
Contact SICK Service.
AI configura-
tion
Number of available and config-
ured analog inputs not identical.
●
No parameters set for AI
●
Connection error
●
Module failure
▸
Check configuration (
log inputs parameters”, page 76
▸
Contact SICK Service.
Interface mod-
ule
No communication via inter-
face module.
●
No parameters set for
module
●
Connection error
●
Module failure
▸
Check configuration (
MCUDH Ex-3K control unit”, page 66
) or
(
see “Connecting to MCUDH Ex-3K control
▸
Contact SICK Service.
No sensor
found
Sender/receiver unit not recog-
nized
●
Communication problems
on RS485 line
●
Supply voltage problems
▸
Check system settings.
▸
Check connection line.
▸
Check voltage supply.
▸
Contact SICK Service.
Variant config-
uration error
MCU setting does not match the
connected sensor.
Sensor type has been
changed
▸
Correct application settings (
the control unit to the sender/receiver unit”,
page 71
Test mode
enabled
MCU in “Test” mode.
▸
Deactivate “System Test” mode
(“Maintenance” directory)