8026957/AE00/V1-0/2021-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SB30
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
4.2.8
Resetting parameters
The device can be reset to factory settings again after parameter changes. A parameter
change in the meantime can also be restored by an automatic backup, which the device
creates after each device restart.
Range low
Lower Measuring
Range Limit
Physical value at live zero
Range high
Upper Measuring
Range Limit
Physical value at 20 mA
Output check cycle
results
Inactive
Control values (
) are not output on the analog
output.
Activated
Control values (
) are output on the analog
output.
cc2
Quadratic
Entry of the regression factors determined using gravimetric comparison
measurement during a calibration (
see “Calibration for dust concentration
cc1
Linear
cc0
Absolute
Entry field
Parameter
Remark
Entry field
Parameter
Explanation
Set to last backup
Parameters are reset to the status
after the last restart
Modbus
®
Remote Terminal Unit (binär)
Modbus
®
ASCII
When switching from RTU <--> ASCII, the selection of data, parity
and stop bit must be deliberately reset!
Set to factory settings
Device is reset to factory settings
Customer parameter settings are lost and reset to default values,
see “Factory settings”, page 36
ResetParaUser
Execution of the selected backup
when pressed
After pressing, the device is reset, the device restarts and SOPAS
should be reconnected to the device or all parameters must be
read in again.