O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SB30
Subject to change without notice
START-UP AND PARAMETER SETTINGS
4
4.2.7
Factory settings
The steps required to modify these settings are described in the following Sections. For this
purpose, the devices must be connected in SOPAS ET (
see “Connection to the device via
), the Level 1 password set and the “Maintenance” mode set.
▸
Connect the measuring system to the SOPAS ET program.
▸
Enter Level 1 password (
see “Password for SOPAS ET menus”, page 40
).
▸
Set the measuring system to “Maintenance” mode.
Fig. 24: SOPAS ET menu: DHSB30/Maintenance/Maintenance
▸
Switch to directory “Application Parameters” and set the desired parameters.
Fig. 25: SOPAS ET menu: Configuration/Application parameter (example)
Entry field
Parameter
Remark
Mounting location
Name of the
measuring location
Sender/receiver unit assignment to the respective measuring location
Response time (T90) Value in s
Response time for the measured variable (
Setting range 1 ... 600 s
Limit value
Value
The limit value relay switches when the value entered is overflown or
underflown.
Relay 3 signals
Maintenance
The relay switched on when “Maintenance” mode is set.
Function check
The relay switches on during a function check procedure.
Maintenance request The relay switches on for “Maintenance” mode.
Interval function
check
Time between two
check cycles
Live zero
Zero point
(2 or 4 mA)
Select 2 or 4 mA to ensure being able to differentiate between measured
value and switched off device or interrupted current loop.