8.5.4
Malfunctions in reading mode: errors when outputting the result status
Prerequisite
The "Device Ready" LED lights up. If not, see
8.5.5
Malfunctions: configuration errors (parameterization)
Prerequisite
The "Device Ready" LED lights up. If not, see
Malfunction
Possible cause
Remedy
1.
The "Result 1" (default setting:
"Device Ready") and "Result 2"
(default setting: "Good read")
switching outputs are not outputting
any pulses.
•
The event of the assigned function for
outputting the result status does not
occur during the reading process
•
Incorrect switching mode parameterized
in CLV
•
not applicable
•
In the CLV Setup program: Check by
choosing the D
EVICE
C
ONFIGURATION
tab.
Click the R
ESULT
O
UTPUT
P
ARAMETERS
button. Are the outputs set to "Invert
Output"?
Change if necessary. Download to CLV!
2.
The beeper is not confirming the
output of the reading result (default
setting: "Good read").
Beeper deactivated
In the CLV Setup program: Check: Check by
choosing D
EVICE
C
ONFIGURATION
tab, group
B
EEPER
V
OLUME
: is "Low" or "High"
selected?
Table 8-6: Troubleshooting: errors in the result status output in reading mode
Malfunction
Possible cause
Remedy
1.
AutoSetup:
The CLV does not read the presented,
application-specific bar code (no
confirmation from beeper)
•
CLV not in "AutoSetup" mode
•
If AutoSetup is started with profile bar
code no. 10: wait time of 5 s after power-
up has been exceeded
•
Bar code presented at incorrect reading
distance
•
Bar code is in total reflection zone
•
Bar code quality not adequate
•
Check: Does the "Device Ready" LED
light up? If so, the CLV is in "reading"
mode again.
In CLV Setup program: Restart
A
UTO
S
ETUP
from the V
IEW
pull-down
menu.
•
Switch the CLV off and on again.
Present profile bar code no. 10 within
5 s.
Then present the application-specific bar
code.
•
Check: Is the reading distance of the bar
code within the specified reading area?
•
Optimize bar code alignment (scan line
should strike the bar code at an angle of
approx. 15
°
).
•
See
Malfunction 1
,
2.
Profile programming:
CLV does not read the presented
profile bar code (no confirmation from
beeper)
•
Wait time of 5 s after power-up has been
exceeded
•
Wait time of 10 s after the first profile bar
code was presented has been exceeded
•
In reading mode: CLV not pulsed
•
Switch the CLV off and on again.
Present profile bar code within 5 s.
•
Once the first profile bar code has been
read successfully, present the next bar
code within 10 s, etc.
•
Clock the CLV accordingly and present
the first profile bar code. Present the next
bar code within 10 s, etc.
Table 8-7: Troubleshooting: configuration errors (parameterization)