Reject level
Sets the range to measure when measuring the peak position of the received
light waveform.
Use the reject level to specify what the lower percentage of the received light
waveform to ignore during calculations.
If you specify a value that is too high, the measured value may be unstable
due to the surface status.
If you specify a value that is too low, diffuse-reflected light and noise will be
measured, which may lead to unstable measured values.
This range is used in the
measurement calculation.
((Level of the received light
waveform) + 1)/128
Figure: Determining the area to use in calculations
7.5
Profile Settings
7.5.1
Profile Extraction Settings
Reject level
Sets the range to measure when measuring the peak position of the received
light waveform.
Use the reject level to specify what the lower percentage of the received light
waveform to ignore during calculations.
If you specify a value that is too high, the measured value may be unstable
due to the surface status.
If you specify a value that is too low, diffuse-reflected light and noise will be
measured, which may lead to unstable measured values.
Target
Select the measurement target from “Normal,” “Gap,” and “Semi-trans.”
“Normal” is standard.
Use this setting to specify what parts of the received light waveform to use
as measurement data.