7 Modes of Operation
7.1
R-L-C-D Mode
This menu item is for measurement of Resistors, Capacitors, Inductances and Diodes. To select the
mode, select R-L-C-D in the main menu. A typical screen looks as follows:
7.1.1
Measurements
When the measured component is connected to the test leads, the screen displays information
depending on the component and the selected settings in the R-L-C-D mode menu.
Capacitance
Inductance
Resistance
Diode
In automatic mode measurement of capacitance larger than 1
μ
F is made at 120 Hz, while ESR is
measured at 100 kHz. In this case the frequency of the test signal is not shown on the screen. For ESR
based rejection of electrolytic capacitors two tables are used, one for standard aluminum electrolytic
capacitors and the other for
Low ESR
capacitors.