quick reference guide

 

DATOSPIR MICRO  

Models 

and

 C

Thank you for purchasing a SIBELMED spirometer. SIBELMED is a leading manufacturer of  
spirometry technology. SIBELMED spirometers are supplied with multiple functions that allow  
absolute control and maximum comfort while performing spirometry maneuvers.

This is a concise guide that will allow you to perform the basic functions quickly. For maximum  
performance, refer to the USER MANUAL

.

Installation / Location

The 

DATOSPIR MICRO

 is totally portable and autonomous and does not require special installation. 

Be careful not to use the equipment in places that:

- Are not safe for the patient or can cause patient stress. 

- Can be subject to splashing of water or other liquids.

- Are subject to strong air currents.

Once the batteries have been placed correctly, the equipment is ready to operate.

Use products that are certified in accordance with directive 93/42/CEE which ensure the patient’s 
safety and the quality of the spirometers.

MY FIRST SPIROMETRY 

step by step

(A) 

Perfoming a FVC (Forced Vital Capacity) maneuver:

1. Start the equipment by pressing               

1 second.

2.   The 

MAIN SCREEN

 appears.

3.  Press the               key to view the screen to enter patient parameters, 

(a numeric or alphanumeric 

screen will appear, depending on the mode selected).

4.  Enter the patient’s parameters and press               A screen containing the axes for plotting the  

maneuver will appear.

5.  After having instructed the patient and the patient is ready for the maneuver, press the               key, 

and wait until a blinking arrow               appears on the screen

.  

THE FVC MANEUVER CAN NOW BE STARTED

w

w

w.sibelmed

.com

w

w

w.sibelmed

.com

Once the maneuver is finished, the values of the 

FVC

 and 

FEV1

  

parameters appear on the top right-hand of the screen, underneath  
the 

REF

erence values (theoretical).

If three or more maneuvers have been performed and the FVC and/or 

FEV1

 parameters blink on and off, this indicates that repeatability  

criterion has been fulfilled for one or both parameters..

This criterion indicates that the two best 

FVC

 or 

FEV1

 values do not differ by more than 150 ml if 

FVC

 is 

more than 1 liter or more than 100 ml if 

FVC

 is less than or equal to one liter. 

The device allows us to perform any number of necessary maneuvers. The 

DATOSPIR MICRO

 always 

stores the five best maneuvers for Forced Vital Capacity 

(FVC)

, and orders them according to 

ATS/ERS 

criteria

, so that 

M1

 is the best and 

M5

 the worst.

Press on 

AXES

 to switch from one graphic 

type

 to the other  

(Flow/Volume or Volume/Time).
Press on 

PARAMETERS

, to view the screen containing the data of the 

selected maneuver.
Press on a 

MANEUVER

, to select it. This allows you to see a graphic, 

consult parameters or print out a report.

Press on the 

GRAPHIC

 area to delete the regular keys and enable the magnifying key and use            or 

             to enlarge or reduce. Press on the same area again to return to the regular keys view.

Press on the 

REFERENCE

 area to access the patient’s data.

To view the second set of keys, press             and to return to the first set of keys, press             
Remember that the             key can be used to return to the menu without losing the available infor-
mation, except if a new patient code is entered or on certain other occasions, but a warning is always 
displayed before loosing the data.

If the warnings are enabled, once the maneuver is finished, one or more 

WARNINGS

 may appear on the screen indicating that the maneuver has 

not been performed according to one or more 

ATS/ERS

 criteria.

FP-

 

This indicates that expiration did not end satisfactorily because  

the variation of accumulated volume during the last second of the  
maneuver was over 25 ml, or that the maneuver lasted for less than  

6 seconds (in individuals 10 or older) or less than 3 seconds (in children under 10).

EX-

 

This indicates that the expiration did not start satisfactorily because the extrapolated volume was greater than 5% of the FVC or 0.15 liters. 

According to  the ATS/ERS recommendations, this volume should be lower than 5% of the FVC or 0.15 liters, whichever is greater.

WARNING:

 The 

DATOSPIR MICRO

 always stores the five best Forced Vital Capacity 

(FVC)

 maneuvers and these  

are ordered according to 

ATS/ERS

 criterion, meaning that 

M1 is the best and M5 the worst

. Therefore, to present 

the diagnosis, print the report or save the maneuver for 

POST

bronchodilation or for the internal DataBase it is  

recommendable to use 

M1, unless the technician deems it convenient to use a different one.

511-

A21-

GR2   R

ev

.  2.0

SIBEL S.A.U.

, Rosellón 500 bajos, 08026 BARCELONA (Spain)

National Sales Office: Tel. 93 436 00 08 - E-mail: [email protected]

International Sales: Tel. +34 93 436 00 07 - E-mail: [email protected] 

After Sales Service: Tel. +34 93 433 54 50 - E-mail: [email protected]

Fax: +34 93 436 16 11     

www.sibelmed.com

 

0318 

Producto clase IIa según Directiva 93/42/CEE