N
Chapter
6
- Taking voltage output signals
Connect external equipment (2k
Ω
or more
load resistance) to + and - of the DS-6200.
- Taking current output signals
Connect external equipment (350
Ω
or less
load resistance) to CUR and G of the DS-
6200.
- Resolution
The D/A converter has a resolution of 1/10000 to -10 - +10V (4 - 20mA).
■
Scale setting value selection
Four types of zero scale and full scale settings can be stored.
Switch with the selector button (
/
) to conduct scale setting.
Scale of output of D/A converter changes accordingly when gain changes by the selection of
calibration value.
How to set
■
Zero scale output value and full scale output value
Setting range:
-
99999 - 99999 (where zero set value < full scale set value)
- In the case of voltage output
Zero scale:
Set the indicated value for the time of output of 0V.
Full scale:
Set the indicated value for the time of output of 10V.
- In the case of current output
Zero scale:
Set the indicated value for the time of output of 4mA.
Full scale:
Set the indicated value for the time of output of 20mA.
How to set
+
-
+
-
F.G
←
Inside Outside
→
DS-6200
External equipment
Load resistance:
2k
Ω
or more
+
-
F.G
DS-6200
+
-
←
Inside Outside
→
External equipment
Load resistance:
350
Ω
or less
CAUTION
- The D/A converter is optional.
- Do not apply external voltage because it will break.
- Do not short-circuit the voltage output because a failure will occur.
Also, connecting a capacitive load may cause oscillation.
SET.
Option Setting (D/A OUTPUT)
Page 1
→
→
System Setting
→
SET.
Option Setting (D/A OUTPUT)
Page 1
→
→
System Setting
→