2
INSTALLATION & CONNECTION
21
INSTA
LLATION &
CON
N
ECTION
Chapter
2
■
Load cell connection
- 4-wire sensor
- 6-wire sensor
Short-cEXC with +S and
-EXC with -S for connecting a
6-wire strain-gage sensor.
■
SI/F connection
The 2-wire serial interface has connective ability for coupling a printer, external display, etc.
Connect from A11 and A12 of the external input/output connector.
The interface is nonpolarized and up to three external instruments may be connected.
A two-core parallel cable or a cabtyre cable (Wire with covering thickened for construction) may be
used for connection.
When a two-core parallel cable or a cabtyre cable is used, the transmitting distance is
approximately 30m. When a two-core shielded twisted pair cable is used, the transmitting distance
is approximately 300m.
Do not parallel it with AC lines and high-voltage lines. It may cause of malfunction.
(
+
EXC)
(
-
EXC)
(
+
SIG)
(
-
SIG)
(SHIELD or
)
+
IN
-
OUT
-
IN
+
OUT
1
2
3
4
DS-6200
5
(
+
EXC)
(
-
EXC)
(
+
SIG)
(
-
SIG)
+
IN
-
OUT
-
IN
+
OUT
DS-6200
1
2
3
4
(SHIELD or
)
5
Converter
Display
Printer
SI/F
A12
A11
←
Inside
Outside
→
DS-6200
Содержание DS-6200