Operation Manual 

Shodex

 IC SI-35 2B 

 

Please read this manual carefully to achieve the best and consistent column performance for a long time

 

 
 

Important Handling Instructions 

 

 Caution! 

 

 

· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column 

and understand their proper handling methods to prevent potential health hazards or death 
from occurring. 

 

· Please wear appropriate personal protective equipment such as lab goggles and gloves 

when handling organic solvents and acid and alkaline reagents. Avoid any direct physical 
contact to prevent chemical injuries.   

 

Before Using the Column 

(1) Please visually inspect the package and outside of the column for any damage.   

 

(2) Please check if product name and serial number written on the column package, column adhesive label, and     

enclosed Certificate of Analysis (CoA) are matching and correct.   

 

1. Introduction 

Thank you for purchasing the Shodex

 product. Shodex

 IC SI-35 2B is a suppressor type ion chromatography 

column for analysis of anions, using sodium carbonate eluents. The column is suitable for the rapid

 analysis of 

standard inorganic anions including fluoride, chloride, nitrite, bromide, nitrate, phosphate, and sulfate. 
 

2. Column Components   
 
 
 

 
 
 

               

 

 

 
 
 
3. Column Specifications 

Product Code 

Product Name 

Column Size (mm) 

Particle Size   

(

μ

m) 

Theoretical Plate 

Number 

(Per Column) 

I.D. Length 

F6995291 

IC SI-35 2B 

2.0 50  3.5 

 4,000 

 

Base Material 

: Spherical porous particles of polyvinyl alcohol modified with quaternary ammonium group

 

Column Housing  : PEEK   
Screw Type 

: Internally-threaded type No.10-32 UNF 

Shipping Solvent : 1.0 mM Sodium Carbonate So 2.0 mM Sodium Bicarbonate Solution 

 

Product Code 

Product Name 

Contents 

F6709720 

IC SI-2GF 

One holder and one filter 

F6709730 

IC SI-2GF filter 

Three filters 

 

Holder Material

 

: PEEK

 

Filter Material

 

: PEEK

 

Filter Pore Size

 

: 2 

m

 

Holder End-screw Type

 

: Internally-threaded type No.10-32 UNF

 

Содержание IC SI-35 2B

Страница 1: ...cate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC SI 35 2B is a suppressor type ion chromatography column for analysis of anions using sodium carbonate eluents The column is suitable for the rapid analysis of standard inorganic anions including fluoride chloride nitrite bromide nitrate phosphate and sulfate 2 Column Components 3 Colu...

Страница 2: ...th a 0 45 μm disposable filter to prevent the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or ha...

Страница 3: ... the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at less than 0 1 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the s...

Страница 4: ...od 1 Insoluble components that block the column inlet may be removed by reversing the flow direction i e introducing the eluent from the column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at 0 1 mL min Method 1 Adsorption of hydrophilic compound...

Страница 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: