Instruction Manual Design and Illustration
6
2 Design and Illustration
2.1 Parameters
2.1.1 Basic parameters
Table 2-1
Technical Index
Parameters
Temperature Control Range
40-105
℃
Heating Rate
≥
1.5
℃
/s
Cooling Rate
≥
1.5
℃
/s
Temperature Accuracy
±
0.5
℃
Temperature Uniformity
±
1
℃
Chip Reaction Volume Range
2 μL
Fluorescence Channel
FAM
Input Voltage
AC100~240V
(
50/60Hz
)
Net Weight
2.5kg
Dimension
330mm
×
250mm
×
180mm
(
exclude packaging
)
Operating Software
PCR Nucleic Acid Analyzer control system 1.0
Hardware version
PCR Nucleic Acid Analyzer hardware system V1.0
Operating System
Linux system
Signal Interface
USB, Ethernet
2.1.2 Functionality index
2.1.2.1 Module operating temperature range
a) Module minimum operating temperature: 40
C, error:
1
C;
b) Module maximum operating temperature: 105
C, error:
1
C.
2.1.2.2 Average heating rate of the module: from 50
C~90
C
,
≥ 1.5
C/s.
2.1.2.3 Average cooling rate of the module: from 50
C~90
C
,
≥ 1.5
C/s.
2.1.2.4 Module temperature control accuracy (temperature control fluctuation): ≤
0.5
C.
2.1.2.5 Module Temperature Accuracy:
Difference between measured temperature and the set temperature ≤
0.5
C.
2.1.2.6 Module temperature uniformity:
Temperature difference ≤
1.0
C.
2.1.2.7 Temperature Duration Accuracy:
The relative deviation of the actual temperature duration from the set temperature duration is ≤
5.0%.
2.1.2.8 Fluorescence intensity detection
a) Repeatability of fluorescence intensity detection (intra-assay measurement repeatability):
Repeatedly detect the high-, medium-, and low-concentrated calibration dye, and the coefficient
of variation (CV) shall be ≤ 3.0%.