60
14. Application Functions
Read Chapter 7 Before Use of
These Pages
14.6 Animal Weighing
Designed for weighing animals. The Animal symbol is illuminated when the Animal Weighing mode is
active.
Select menu item
.
1
Load the weighing vessel and press the
[O/T]
key.
2
Load the sample (animal etc.) with a mass
more than 50 times the Zero Range.
3
When the weighed value is relatively stable, the
value is automatically output.
4
Press the
[O/T]
key or unload the sample.
5
When the displayed value is stable and less
than 10 times the Zero Range, automatic zeroing
occurs.
Any residue remaining on the pan (excrement or
fur) is automatically canceled and zeroing occurs.
If zeroing does not occur, increase the value of
the Zero Range (13.5).
53
Note
Data may be output when the weighing vessel is loaded. This is not a mal-
function.