11.3 Stability Detection Band
The conditions for indicating balance stability can be selected. If “1 count” is selected, when the display
has remained constant (within one display count), the balance is regarded as stable and the stability
mark illuminates. Stability detection band settings can be selected from 2 to 64 counts.
Select menu item
for “1 count”.
for 2 counts,
for 4,
for 8,
for 16,
for 32, and menu
item
to set the band as 64 counts.
11.4 Tracking
Tracking is the function that will maintain the current displayed value as long as possible.
To turn ON this function,
Select menu item
.
To turn OFF this function,
Select menu item
.
Note
Use “16 counts” to “64 counts” only when the Auto Print function is used
and it is therefore required that the sensitivity to vibration be reduced to
allow the balance to be regarded as stable and print the measurement
value.
The balance may not operate properly if these settings are selected under
normal use conditions and environment.
Zero Tracking Function
When the display is zero, tracking functions as “zero tracking” to keep the
zero display as long as possible. “Zero tracking” automatically cancels
small zero drift.
It is recommended to set tracking off when measuring slight mass change
such as in the process of drop addition or liquid evaporation.