34
10. Calibration
Read Chapter 7 Before Use of
These Pages
10.2.3 Span Calibration Using External Weights
1
Verify that the balance is in mass display and
that the pan is empty.
2
Press the
[CAL]
key once. “E-CAL” is dis-
played.
(If “E-CAL” is not displayed, return to mass dis-
play and select menu item
.)
3
Press the
[O/T]
key.
The value of the correct calibration weight to be
loaded is displayed and blinks.
4
Load the indicated calibration weight and press
the
[O/T]
key.
5
Shortly, zero display blinks. Unload the weight
from the pan and press the
[O/T]
key.
“SEt” is displayed briefly to indicate completion of
span calibration.
(Example)
3
Changing the Calibration Weight to be Used
Pressing the
[CAL]
key allows changes to the weight value. Modify the
value using the
[UNIT]
key and
[PRINT]
key, then press the
[O/T]
key. To
interrupt modification, press the
[POWER]
key.
Note
OIML Class E2 or F1 calibration weight is recommended for calibration,
depending on your accuracy demand.