4. Installation
Read These
Pages
Before
Installation
UX series [Span Calibration Using External
Weights]
1
Verify that the balance is in gram-display and
unload the sample from the pan.
2
Press the
[CAL]
key once. “E-CAL” is displayed.
3
Press the
[O/T]
key.
The value of the correct calibration weight to be
loaded is displayed and blinks.
4
Load the indicated calibration weight and press
the
[O/T]
key.
5
When the zero display blinks, unload the weight
from the pan and press the
[O/T]
key. “Set” is dis-
played briefly to indicate completion of span cali-
bration. Then the gram-display will return.
(Example)
Note
Span calibration is required again :
when the location of the balance is changed,
when the room temperature changes considerably,
periodically, according to the quality control plan of the user.