4. Installation
Read These
Pages Before Installation
4.5 Span Calibration
It is necessary to calibrate the balance after it is moved.
Verify that the balance is stable before performing the span calibration. To achieve a very stable state,
ensure that the balance has been turned on with the gram-display for at least one hour, that the tem-
perature is constant, that there are no breezes or vibrations and that the balance is in an area isolated
from the normal traffic flow.
UW series [Span Calibration Using the Built-
in Weight]
1
Verify that the balance is in gram-display and
that the pan is empty.
2
Press the
[CAL]
key once. “i-CAL” is displayed.
3
Press the
[O/T]
key. After “i-CAL3”...“i-CAL1”,
“Set”, “CALEnd” are displayed indicating the com-
pletion of span calibration, the gram-display will
appear.
This is the standard calibration type. Refer to
for use of external weights.
Note
Span calibration is required again :
when the location of the balance is changed,
when the room temperature changes considerably,
periodically, according to the quality control plan of the user.