4 Operation
4-14
Note
After master curve correction, carry out check analysis using the
sample for checking (below).
4.9
Check analysis
Confirm whether the content percentages of the elements are correctly measured,
using a sample with known composition.
1 Select
Analysis
-
Cont.
on the menu bar.
2 Input the name of the analytical group in the
Group
textbox or select it by
pressing
S4:Group
key.
Screen 4.5 Content analysis
3 Set the sample on the spark stand and hold it by the sample holder.
4 Start the discharge by pressing
1:Start
key or the start button on the spark stand.
5 Take out the sample, and change the position of the sample to be analyzed.
Then, repeat procedures 3 and 4 specified times.
6 Confirm that the correct content percentages are obtained.