Chapter 5 Maintenance and Inspection
Fig. 5.2 Remove the Top Cover
Fig. 5.3 After Removing the Top Cover
5.1.1.3 Power Spectrum Check
FTIR performance is evaluated by measuring the power spectrum. The power spectrum gives the
intensity at various wavenumbers. After ensuring that the sample chamber is empty, enter the
following parameters in the Parameter Tab on the [Measurement] Tab Screen, and measure the
power spectrum to verify the specifications.
In [Data] tab:
"Measuring mode":
"Power"
"Resolution":
"2 cm
−
1
"
"Range":
"4700-340 cm
−
1
"
"N scans":
"45" scans
"Apodization":
"HAPP-GENZEL"
"Detector":
"standard"
"Beam":
"internal"
In [More] tab:
"Detector speed":
"2.8 mm/sec"
Lift up to remove the Top cover.
Silica gel case
Confirm the color of the
Silica gel inside
Содержание IRAffinity-1
Страница 2: ...This page is intentionally left blank...
Страница 14: ...xii IRAffinity 1 Symbol Definition Current AC Symbols Found on the IRAffinity 1...
Страница 39: ...2 1 IRAffinity 1 Chapter 2 Specifications CONTENTS 2 1 IRAffinity 1 Specifications 2 2...
Страница 60: ...Chapter 3 Installation 3 18 This page is intentionally left blank IRAffinity 1...
Страница 128: ...Chapter 7 Materials for Infrared Analysis 7 28 This page is intentionally left blank IRAffinity 1...
Страница 131: ...3 IRAffinity 1 Z ZnSe 7 2 7 4...
Страница 132: ...4 This page is intentionally left blank IRAffinity 1...
Страница 134: ...This page is intentionally left blank IRPrestige 21...