4.2 Initialization of IRAffinity-1 and Self Diagnostics
4-3
IRAffinity-1
Initialize the IRAffinity-1 by IRsolution software.
1. Confirm that there is nothing interrupting the IR beam in the compartment of the IRAffinity-1.
If there is a sample or an accessory which interrupts the IR beam, "Infrared power spectrum shape" and
another tests in Self diagnostics fails.
2. Select [Measurement]-[Initialize].
3. Self diagnostics is executed and its result is displayed on the log window.
4. Failed item is indicated as "Failure" on the log window.
5. After all items are inspected, a message "INIT SUCCESS" is recorded on the log window, the condition of
IRAffinity-1 is displayed on the status window, and then IRAffinity-1 is ready to operate.
When "Initialize FTIR on startup" on [Environment]-[Instrument Preferences]-[FTIR] is checked, the software
initializes the IRAffinity-1 automatically upon startup.
NOTE
When the IRsolution software is already activated, IRAffinity-1 should be initialized after starting-up
initialization of IRAffinity-1 finishes (about 20 seconds after activation).
4.2
Initialization of IRAffinity-1 and Self
Diagnostics
Содержание IRAffinity-1
Страница 2: ...This page is intentionally left blank...
Страница 14: ...xii IRAffinity 1 Symbol Definition Current AC Symbols Found on the IRAffinity 1...
Страница 39: ...2 1 IRAffinity 1 Chapter 2 Specifications CONTENTS 2 1 IRAffinity 1 Specifications 2 2...
Страница 60: ...Chapter 3 Installation 3 18 This page is intentionally left blank IRAffinity 1...
Страница 128: ...Chapter 7 Materials for Infrared Analysis 7 28 This page is intentionally left blank IRAffinity 1...
Страница 131: ...3 IRAffinity 1 Z ZnSe 7 2 7 4...
Страница 132: ...4 This page is intentionally left blank IRAffinity 1...
Страница 134: ...This page is intentionally left blank IRPrestige 21...