18 Error Messages
18.1 Error Messages
286
GC-2014
18.1.2
Temperature control errors
Heat loss
This message appears when heat loss is substantial. The oven door may be open or the
insulation has a hole. If the column oven door is open, close it and select “Reset Error.” If the
insulation has a hole in it, turn off the power of the unit, repair the hole, and then restart the
system.
This error may not occur even when the door is open when the column oven temperature is
set at a certain range or the temperature around the unit is a certain level.
Temperature increase limit over
These messages appear when the maximum temperature limit has been exceeded. Press
“Ignore Error” and change the maximum limit temperature. If the confirmation screen to reset
the error appears again, press “Reset Error.”
Code
Message
CS
1001
Heat is escaping
Yes
Code
Message
CS
1002
COL temp exceed the limit
Yes
1003
INJ1 temp exceed the limit
Yes
1004
DET1 temp exceed the limit
Yes
1005
INJ2 temp exceed the limit
Yes
1006
DET2 temp exceed the limit
Yes
1007
AUX3 temp exceed the limit
Yes
1008
AUX4 temp exceed the limit
Yes
1009
AUX5 temp exceed the limit
Yes
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...