16 Special Functions
16.2 Batch Schedule
222
GC-2014
16.2.5
Parameter list
START VIAL NO.
Range:
1
Default: 0
FINAL VIAL NO.
Range:
1
Default: 0
1: With no auto sampler
Short rack ... 1
−
6
Long rack ... 1
−
12
With auto sampler 1
−
150 (maximum)
When the auto sampler is used, the possible depends on the vial rack type and the number
of racks.
The default, “0”, indicates that the setting is not given.
SAMPLE
Inject sample
The specified sample vial No. is analysed.
Without sample
Vials are not used for samples. Instead samples are introduced from a gas sampler with a
valve.
GC FILE NO.
Range: FILE 0
−
9, Default: FILE0
Set the file No. of the GC method to be used.
CHANNEL NO.
Range: Ch1/Ch2, Default: Ch 1
Set the channel No. connecting the GC to the Chromatopac. This setting is ignored when
using GC solution software.
DATA PROCESSOR FILE NO.
Range: FILE 0
−
9, Default: FILE0
Specify a date processor (Chromatopac) date integration file.
−
This setting is ignored when
using GC solution software.
NUMBER OF INJECTION
Range: 0
−
99, Default: 0
Set the number of times to inject each sample.
SAMPLE TYPE
Unknown
Select this item to analyze a sample of unknown concentration.
Analyze a standard and create a calibration curve before analyzing an unknown
sample.
Standard
Select this item to analyze a standard of known concentration.
SLOPE TEST
Range: On/Off, Default: Off
When “On” is selected, the slope test is performed before running the batch schedule.
CLEAN UP
Range: On/Off, Default: Off
When “On” is selected, clean up is performed before running the batch schedule.
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...