10 Starting and Stopping Analysis
10.1 Making an Injecting and Starting an Analysis
111
GC-2014
Q
Injection
10.1.3
Starting the analysis
Start an analysis program when the STATUS light is green.
Press the [START] key to start the temperature program, the time program, the pressure
program and the flow rate program.
Any pre-set programs execute prior to the start of the analysis programs.
Aspirate sample.
Insert syringe until it
touches needle guide.
Push plunger to inject
sample.
At same time, press
[START] key of gas
chromatograph.
Promptly pull out
syringe.
WARNING
Wear protective goggles when handling samples.
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...