INSTRUCTION MANUAL

FOR

3-in-1 Combo Electrotherapy Device

Shenzhen Roundwhale Technology Co., Ltd.

This manual is valid for the R-C1 Stimulator

Model: R-C1

Be sure to read this instruction manual before operate and 

keep it where safe.

This user manual is published by Shenzhen Roundwhale 

Technology Co., LTD.

Содержание R-C1

Страница 1: ...y Device Shenzhen Roundwhale Technology Co Ltd This manual is valid for the R C1 Stimulator Model R C1 Be sure to read this instruction manual before operate and keep it where safe This user manual is published by Shenzhen Roundwhale Technology Co LTD ...

Страница 2: ...Amendments will be published in a new edition of this manual All Rights Reserved R C1 Rev V1 1 2018 printed in Sep 11 2018 Declaration of conformity Shenzhen Roundwhale Technology Co Ltd declares that the device complies with following normative documents IEC60601 1 IEC60601 1 2 IEC60601 1 11 IEC60601 2 10 IEC62304 ISO10993 5 ISO10993 10 ISO10993 1 ISO 14971 ...

Страница 3: ... 4 SPECIFICATION 13 5 OPERATING INSTRUCTION 14 6 INSTRUCTIONS FOR USE 20 7 Program 26 8 Cleaning and maintenance 28 9 Troubleshooting 30 10 Storage 31 11 Disposal 32 12 Electromagnetic compatibility EMC tables 32 13 Normalized symbols 37 14 WARRANTY 38 TABLE OF CONTENS ...

Страница 4: ...eir intensity frequency and pulse width The parameters for each program are list ed in clause 7 1 Based on simulating the body s natural pulses the mech anism of electrical stimulation equipment is to create electric impulses that are transcutaneous transmitted to nerves or muscle fibers through electrode The intensity of the dual channel can be adjusted independently and can be applied individual...

Страница 5: ... TENS TENS Transcutaneous Electrical Nerve Stimulation is ef fective in relief of pain It is daily used and clinically proven by physiotherapists caregivers and top athletes around the world High frequency TENS currents activates the pain in hibiting mechanisms of the nervous system Electrical im pulses from electrodes placed on the skin over or near the pain area stimulate the nerves to block the...

Страница 6: ...S stimulation program stimulates healthy muscles in order to improve and facilitate muscle performance Massage mode The Massage stimulation program provides relaxing muscle vibration to loosen tight muscles 2 2 Important Safety Precautions and Warnings It is important that you read all the warnings and precautions included in this manual because they are intended to keep you safe prevent risk of i...

Страница 7: ...ter the operation 8 Do not use with serious arterial circulatory problems in the lower limbs 2 2 2 WARNING 1 If you have had medical or physical treatment for your pain consult with your physician before use 2 If your pain is not subdued becomes more than mild or lasts for more than five days stop using the device and consult with your physician 3 Do not apply stimulation over your neck because th...

Страница 8: ...r is con nected to high frequency surgical equipment which may cause burn injuries on the skin under the elec trodes as well as problems with the stimulator 13 Do not use the stimulator in the vicinity of shortwave or microwave therapy equipment since this may affect the output power of the stimulator 14 Never use it near the cardiac area Stimu lation electrodes should never be placed anywhere on ...

Страница 9: ...actitioner qualified in the management of pain patients 5 Since the effects of stimulation of the brain are un known stimulation should not be applied across your head and electrodes should not be placed on opposite sides of your head 6 The safety of electrical stimulation during pregnancy has not been established 7 You may experience skin irritation or hypersensitivity due to the electrical stimu...

Страница 10: ...ch sends electrical pulses to your body 17 Do not use sharp objects such as a pencil or ballpoint tip to operate the buttons on the control panel 18 Check the electrode connections before each use 19 Electrical stimulators should be used only with the electrodes recommended for use by the manufacturer 2 2 4 Adverse Reactions 1 Possible skin irritation or electrode burn under the electrodes may occ...

Страница 11: ...escription 1 Treatment mode 10 Timer sign 2 Treatment body part 11 Standby mode program num ber Working mode treatment time 3 Low battery indicator 12 Key locking symbol 4 SET Sign 13 intensity of Channel2 5 PAUSE Sign 14 Unit of treatment intensity for Channel 2 6 Symbol of Channel 1 15 AUTU Sign 7 Indicator of no load Channel 1 16 Symbol of Channel 2 8 intensity of Channel1 17 Indicator of no lo...

Страница 12: ...atment At query mode press the ON OFF button to standby mode 3 M button At standby mode press the M button to select the treat ment mode 4 B button At standby mode press the B button to select the treat ment body part 5 P button At standby mode press the P button to select the treat ment program 6 Q button At standby mode press the Q button to query the treat ment records 7 button At standby mode ...

Страница 13: ...able Treatment mode TENS EMS and MASSAGE mode Output characteristics Constant Current CC Operating condition 5 C to 40 C with a relative humidity of 15 93 atmospheric pressure from 700 hPa to 1060 hPa Storage condition 10 C to 55 C with a relative humidity of 10 95 atmospheric pressure from 700 hPa to 1060 hPa Dimension 120 5x69 5x27 mm L x W x T Weight about 104g without battery Automatic shutoff...

Страница 14: ... P W pulse width 30 220μs P R Frequency 25 290Hz Hz vibration per second Treatment time 30 minutes 5 OPERATING INSTRUCTION 5 1 Battery 5 1 1 Check replace batteries Open the battery cover insert four batteries type AAA into the battery compartment Make sure you are installing the batteries properly Be sure to place the batteries according to the markings of positive end and negative end in the bat...

Страница 15: ... you don t use it for a long time This can prevent device damage due to the battery leakage 5 Replace all of the batteries simultaneously 6 Always replace the device with the same type battery 5 2 Connect electrode pads to electrode wires Insert the electrode wires connector into electrode con nector Make sure they are properly con nected to ensure the good performance Please refer to the picture ...

Страница 16: ...rode options The electrodes are disposable and should be routinely re placed when they start to lose their adhesiveness If you are unsure of your electrode adhesive properties please order new replacement electrodes Replacement electrodes should be re ordered under the advice of your physician or the device manufacturer to ensure proper quality Follow application procedures outlined on electrode p...

Страница 17: ... first in order to avoid un wanted irritation 5 It is recommended that at minimum 1 97 x 1 97 self adhesive square electrodes are used at the treat ment area 6 Never remove the self adhesive electrodes from the skin while the device is still on 5 4 3 Electrode placement R C1 is a kind of OTC stimulator suitable for home use You only have to use according to the user manual place the electrode on t...

Страница 18: ...18 Back Arm Top of hand Palm Leg Joint ...

Страница 19: ...19 Joint Hip Position of electrode placement under EMS programs Arm Back Leg Shoulder Hip ...

Страница 20: ... button to turn the device on the LCD will be lit And then it goes into the standby mode as below picture shows 6 2 Select treatment mode Press the M button to select which treatment mode TENS MASS EMS you will use The LCD displays as follows 6 3 Select treatment body part Based on your needs press B button to select the current treatment body part Press each B button to switch body part The LCD d...

Страница 21: ... intensity The LCD displays as follows 6 6 Adjust the output intensity Place the electrodes on the body parts press the button to increase output intensity It will be increased to a higher level after each press The device has totally 60 levels of output intensity Please adjust the intensity to the condi tion that you feel comfortable The level of output intensity will be shown on the LCD If you f...

Страница 22: ... and consult with your medical practitioner if problems insist 6 7 Pause the treatment On the treating mode press the Q button to pause the treatment and the LCD will display the icon Press the Q button again to regain the treatment 6 8 Stop the treatment and turn off the device Press the ON OFF button to stop treatment during the treating mode Press the ON OFF button again to turn off the stimula...

Страница 23: ...es And the stimulator returns to the standby mode 6 10 Query the treatment records If you want to query the treatment records press the Q button to check the details The LCD will show as follows Press the button or button to view the treatment his tory Press the on off button to return to the standby mode 6 11 Low battery detection When the battery is low the icon will twinkle to indi cate it You ...

Страница 24: ...he sound prompt function is on press and hold the B button and the device will emit a long DI to turn off the sound prompt function b In the waiting state when the sound prompt function is off press and hold the B button and the device will make two short DI to turn on the sound prompt function Note Reinstall the battery and the sound prompt func tion is turned on 6 13 Usage of electrode pads a Th...

Страница 25: ... Before applying the electrode it is recommended for us ers to wash and degrease the skin and then dry it 2 Never remove the electrode from the skin while the de vice is still on 3 Only use the electrode pads provided by the manufac turer Use of other companies products could result in injuries to the user 6 14 Where do I attach electrode pads a Each person reacts differently to electric nerve sti...

Страница 26: ...period of treatment if the pain relief is not achieved or the pain gets even worse please consult your doctor Usage advice for EMS a Place the electrodes on the body part you want to treat referring to the picture on Section 5 4 3 b 1 2 treatment per day about one week as a period of treatment c We advise you use the device for one session per time If you feel discomfort during treatment you can e...

Страница 27: ...50 30 P2 Burst 100 150 30 P3 FM 6 8 10 250 30 Joint P1 Han 2 100 200 150 30 P2 Normal 100 150 30 P3 Burst 80 150 30 In EMS mode Applica ble parts Pro gram Frequency Hz Pulse width uS Treatment time min Prepare stage Work stage Relax stage Shoul der P1 5 55 6 3 280 28 P2 5 75 4 3 280 32 P3 5 110 1 3 280 30 Back P1 5 55 6 3 300 28 P2 5 75 4 3 300 32 P3 5 110 1 3 300 30 Arm P1 5 55 6 3 200 28 P2 5 75...

Страница 28: ...th the following necessary daily mainte nance requirements to make sure the device is intact and guarantee its long term performance and safety 8 1 Cleaning and care for the device 8 1 1Pull the electrodes out of the stimulator remove the batteries and clean the device with a soft slightly damp cloth In case of heavier dirt build up you may also apply a mild detergent 8 1 2 Do not expose the COMBO...

Страница 29: ...ry is unloaded before cleaning 8 2 Maintenance 8 2 1 The manufacturer didn t authorize any maintenance agencies abroad If your device has any problem please contact the distributor The manufacturer will not be responsible for the results of maintenance or repairs by unauthorized persons 8 3 2 The user must not attempt any repairs to the device or any of its accessories Please contact the retailer ...

Страница 30: ...nter face 4 The battery is not the right model or some thing goes wrong with the battery interface 5 Exception reset 1 Check and clean the compartment 2 Replace the new battery or install the battery correctly 3 Check and clean the interface 4 Replace the battery with the right model No sensation of stimulation or weak stimulation 1 The electrode does not connect well to the skin 2 If the connect ...

Страница 31: ... well 3 Wipe the electrode with a wet cotton cloth before use 4 Check your aller gic history Please change the sticking place or shorten the treatment time If your skin is over sen sitive you should stop the treatment or go to see a doctor 10 Storage Remove the batteries from the device if you are not going to use it for a long period Battery leakage can damage the unit Do not make any sharp pull ...

Страница 32: ...ipment may have potentially harmful effects on the environment Incorrect disposal can cause toxins to build up in the air water and soil and jeopardize human health 12 Electromagnetic compatibility EMC tables Guidance and manufacture s declaration electromagnetic emis sions The device is intended for use in the electromagnetic environment specified below The customer or the user has to assure that...

Страница 33: ...ow The customer or the user of the device should assure that it is used in such environment Immunity test IEC 60601 Test level Compliance level Electromagnetic environment guid ance Electrostatic discharge ESD lEC61000 4 2 8kV direct indirect contact 15kV air discharge 8kV direct indirect contact 15kV air discharge Floorsshouldbe wood concreteor ceramictile Iffloors arecoveredwith syntheticmateria...

Страница 34: ... 61000 4 8 10V m 10V m Powerfrequencymag neticfieldsshouldbe atlevelscharacteristic ofatypicallocationin atypicalcommercial orhospitalenviron ment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacture s declaration electromagnetic im munity The device is intended for use in the electromagnetic environment specified below The customer or the user of device...

Страница 35: ...wer rating of the transmitter in watts W according to the trans mitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicin ity of equipment marked with the following symbol NOTE 1 At 80 ...

Страница 36: ...d additional measures may be necessary such as reorienting or relocating the Blood Pressure Monitor b Over the frequency range 150 kHz to 80 MHz field strengths should be less than Vi V m Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Table 9 Test fre quency MHz Band a MHz Servicea Modula tion b Maxi mum power W Dis tance m Immu nity Test Level V m 385 380 ...

Страница 37: ...cies are in cluded b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alternative to FM modulation 50 pulse mod ulation at 18 Hz may be used because it does not rep resents actual modulation It would be worst case 13 Normalized symbols Electrical devices are recyclable material and should not be disposed of with household waste after use Help us to protect the enviro...

Страница 38: ...e device centre in case of a claim under the warranty If you have to return the unit en close a copy of your receipt with clear statement of defect description The warranty terms is below 1 The warranty period for this device is 1 year from date of purchase In case of a warranty claim the date of pur chase has to be proven by means of the sales receipt or invoice 2 Repairs under warranty should be...

Страница 39: ...turer to the consumer or the service centre Accessories which are subject to normal wear and tear Device damage due to privately dissembling devices 4 Liability for direct or indirect consequential losses caused by the unit is excluded even if the damage to the unit is accepted as a warranty claim ...

Страница 40: ...hen Roundwhale Technology Co Ltd Add 202 2 F Building 27 Dafa Industrial Park longxi community longgang street longgang district Shenzhen China Tel 86 755 23212776 Fax 86 755 23212776 E mail info roovjoy com ...