shelolab SMO14-2 Скачать руководство пользователя страница 1

 

Installation - Operation Manual 

SMO14-2 SMO28-2 SMO28G-2 

SMO FORCED AIR OVENS 

230 Voltage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание SMO14-2

Страница 1: ...Installation Operation Manual SMO14 2 SMO28 2 SMO28G 2 SMO FORCED AIR OVENS 230 Voltage...

Страница 2: ...2 P a g e These ovens require permanent connect wiring also known as a hardwiring to a power source Pictured on Cover Left to right SMO14 2 and SMO28 2...

Страница 3: ...rtifications These units are CUE listed by T V S D as forced air ovens for professional industrial or educational use where the preparation or testing of materials is done at an ambient air pressure r...

Страница 4: ...ring 15 Lifting and Handling 16 Leveling 16 Install the Oven 17 Installation Cleaning 17 Install the Shelving 17 GRAPHIC SYMBOLS 19 CONTROL PANEL OVERVIEW 21 OPERATION 23 Operating Precautions 23 Theo...

Страница 5: ...rs must draw up a site specific standard operating procedure SOP covering each application and associated safety guidelines This SOP must be written and available to all operators in a language they u...

Страница 6: ...brated by a third party For best results use a digital device with one or more thermocouple probes Remote readings eliminate chamber door openings and avoid subsequent waits for the chamber temperatur...

Страница 7: ...find any damage to the unit follow the carrier s procedure for claiming damage or loss 1 Carefully inspect the shipping carton for damage 2 Report any damage to the carrier service that delivered the...

Страница 8: ...Permanent Connect Wire Braid 10 gauge 6 inches 15 cm Control Panel Vent Port with Sliding Dampener Vent Port with Sliding Dampener Main Temperature and OTL Sensor Door Latch Door Latch Shelf Standard...

Страница 9: ...Permanent Connect Wire Braid 10 gauge 6 inches 15 cm Control Panel Vent Port with Sliding Dampener Vent port with Sliding Dampener Main Temperature and OTL Sensor Probes Door Latch Shelf Standard Rail...

Страница 10: ...ain Temperature and OTL Sensor Probes Shelf Standard Rail Chamber Door Handle Vent Port with Sliding Dampener Vent Port with Sliding Dampener Oven Chamber Power Panel Back Permanent Connect Wire Braid...

Страница 11: ...ata plate on the back of the oven below the power braid inlet The data plate contains the oven model number and serial number Technical support and your distributor will need these numbers in order to...

Страница 12: ...12 P a g e RECEIVING YOUR UNIT...

Страница 13: ...al Pre Installation Check that the required ambient condition for the oven are met page 14 Check that the spacing clearance requirements are met page 14 Unit dimensions may be found on page 39 Check f...

Страница 14: ...y and meet its stated temperature specifications 24 inches 610 mm of vertical headspace clearance Do not place objects on top of the oven Allow at least 6 inches 152 mm from the access port and fan ve...

Страница 15: ...uilding installation to protect against overcurrent conditions The required circuit breakers are SMO14 2 15A and SMO28 2 SMO28G 2 30A Power feed disconnect The oven must be positioned so that all oper...

Страница 16: ...abilization Restrain the oven completely while lifting or transporting so it cannot tip Remove all moving parts such as shelves and trays and lock doors in the closed position during transfers to prev...

Страница 17: ...ot clean with deionized water See the Cleaning and Disinfecting topic in the User Maintenance section see page 33 for more information on how to clean the oven chamber prior to putting the unit into o...

Страница 18: ...18 P a g e INSTALLATION...

Страница 19: ...t system Indique le syst me de d passement de temperature AC Power Rep re le courant alternatif I ON O OFF I indique que l interrupteur est en position marche O indique que le commutateur est en posit...

Страница 20: ...C SYMBOLS Symbol Definition Indicates the timer Indique le minuterie Start or Stop the Timer Lancer ou arr ter le minuteur Reset the Timer R initialisation de la Minuterie Caution hot surface Attentio...

Страница 21: ...in the event of a failure of the main temperature controller system For more details please see the explanation of the Over Temperature Limit System on page 25 in the Theory of Operation entry OTL Lig...

Страница 22: ...l as an adjustable offset while in calibration mode The arrow buttons can be used to adjust the temperature set point or place the unit in its calibration mode and then enter a calibration offset valu...

Страница 23: ...s or fuel air mixtures in quantities that can become potentially explosive Outgassed byproducts may be hazardous to or noxious for operating personnel Exhaust should be vented to a location outside th...

Страница 24: ...s powered the chamber air temperature cannot go below the ambient room temperature plus the internal waste heat of the oven Waste heat is generated primarily by the operation of the blower fan motor a...

Страница 25: ...r Temperature Limit System OTL When set the OTL heating cutoff system prevents runaway heating in the event of a failure of the microprocessor controller or its thermometer probe The OTL operates inde...

Страница 26: ...N INTO OPERATION Carry out the following steps and procedures to put the oven into operation after installing it in a new workspace environment 1 Turn on the oven by placing the green main power switc...

Страница 27: ...and hold either Set Temperature Set Point Adjustment Mode Note The display will automatically exit the adjustment mode after 5 seconds of inactivity on the arrow keys saving the last shown set point...

Страница 28: ...y impacting the unit temperature stability If you can find no sources of external or internal temperature fluctuations contact Tech Support or your distributor for assistance End of Procedure 1 Set OT...

Страница 29: ...n The oven will cease heating 2 Place the Timer Display in its adjustable Set Timer mode Press and Hold RESET Set Timer Hours Selected After approximately 3 seconds a flashing decimal point will appea...

Страница 30: ...er Push Set Timer Minutes Selected The flashing decimal point will advance to between the third and fourth numbers saving the new Tens of Minutes parameter setting 7 Set the Minutes parameter Adjust S...

Страница 31: ...viously programmed profile duration The oven will cease heating 2 Launch the current profile Press and Hold Set Timer The Timer Activated light illuminates and the Display will start counting down The...

Страница 32: ...r your application and will not suffer damage when brought to temperature Always set the OTL cutoff system to approximately 5 C above your application temperature set point in order to safeguard acces...

Страница 33: ...rine based bleaches or abrasives these will damage the chamber liner Warning Exercise caution if cleaning the unit with alcohol or flammable cleaners Always allow the unit to cool down to room tempera...

Страница 34: ...items from the chamber when disinfecting Disinfect all surfaces in the chamber making sure to thoroughly disinfect the corners Exercise care to avoid damaging the sensor probes When disinfecting exte...

Страница 35: ...t port not being used 3 The oven chamber door must be closed and latched 2 Place the sesnor probes in the oven with the probe heads at least at least 2 inches 5cm from the surface of the shelving or w...

Страница 36: ...ice and chamber temperature display readings a If the readings are the same or the difference between the two falls within the acceptable range of your protocol the display is accurately showing the c...

Страница 37: ...he new current chamber temperature value Set Temperature Set Temperature Current Temp Value 4 Adjust the current temperature value to match the reference device a Use the UP and DOWN arrow buttons Set...

Страница 38: ...display temperature readings a If the readings are the same or the difference between the two falls within the acceptable range of your protocol the oven is calibrated for temperature Or b Advance to...

Страница 39: ...ries We reserve the right to alter technical specifications at all times WEIGHT Model Shipping Net Weight SMO14 2 361 lbs 164 kgs 280 0 lbs 127 3 kgs SMO28 2 513 lbs 233 kgs 390 0 lbs 177 3 kgs SMO28G...

Страница 40: ...7 2 kg 30lbs with the weight evenly distributed across the shelf AIR FLOW PERFORMANCE Ventilation Rates Model Cubic Feet per Minute 80 C Liters per Minute 80 C SMO14 2 42 1189 SMO28 2 39 4 1116 SMO28G...

Страница 41: ...60 C 3 5 150 C 0 2 C SMO28G 2 Ambient 15 to 260 C 3 5 150 C 0 2 C Model Heat Up Time to 150 Recovery at 150 C SMO14 2 35 minutes 7 minutes SMO28 2 29 minutes 7 minutes SMO28G 2 29 minutes 7 minutes Fr...

Страница 42: ...42 P a g e UNIT SPECIFICATIONS...

Страница 43: ...may order it directly from Sheldon Manufacturing by calling 1 800 322 4897 extension 3 If you are uncertain that you have the correct Part Number or if you need that specific item please contact Sheld...

Страница 44: ...P O Box 627 Cornelius OR 97113 USA support sheldonmfg com sheldonmanufacturing com 1 800 322 4897 503 640 3000 FAX 503 640 1366...

Отзывы: