7
CONTROL PANEL OVERVIEW
4.1
Power Switch:
The main power I/O (on/off) switch controls all power to the unit
and must be in the I or ON position before any systems are operational.
4.2
Temperature Controller:
The controller is marked SET TEMPERATURE and is
equipped with an adjustment knob and a graduated dial. The graduated dial is
marked with 10 major increments and 50 minor increments. These increments
can be used as index points for setting and returning to set point temperatures.
4.3
Heating Lamp:
This green pilot lamp is marked HEATING ACTIVATED. It
indicates when the element has been activated and the incubator is heating.
When set point is reached the pilot light will cycle on and off as the elements
maintain set point.
4.4
Circuit Breaker / Fuse:
This control is adjacent to the power cord and provides
protection for the unit
’s electrical circuitry against power fluctuations. The circuit
breaker, when tripped, must be reset by pushing in the extended button for the
unit to continue operation. The fuse, when blown (on CE units in place of the
circuit breaker) must be replaced before the unit can continue operation.
Section