MECHANICAL 

Door not sealing 

 

 

1/ check physical condition of gasket 
2/ assure that gasket is in original location 
3/ Verify that door latch is operating correctly and not miss-aligned or 
broken. 

OTHER 

Unit or wall fuse/circuit breaker is blown 

 

 

1/ check wall power source. 
2/ compare current draw and compare to specs on data plate.  
3/ see what other loads are on the wall circuit. 
4/ has the circuit breaker on the front panel been tripped? If so, see 
section 4.4 

Unit will not turn on 

 

 

1/ check wall power source 
2/ check fuse/circuit breaker on unit or in wall 
3/ check all wiring connections, esp. around the on/off switch   

Unit is smoking 

– out of box 

 

 

This is not an uncommon occurrence when the protective coatings heat 
up during initial operation.  Put unit under vent and run at full power for 
one hour until the smoke dissipates. 

Contamination in chamber 

 

 

1/ see cleaning procedure in section 6.1 
2/ develop and follow standard operating procedure for specific 
application; include definition of cleaning technique and maintenance 
schedule. 

If  following  these  troubleshooting  suggestions  does  not  solve  the  problem,  call  Customer  Service  for 
assistance.  Please see the manual cover for information on where to contact Customer Service. 

Содержание SMI1E

Страница 1: ...ly Designated as EI1 E1 2 EI2 E2 2 11 2013 4861571 INSTALLATION AND OPERATIONAL MANUAL Sheldon Manufacturing Inc P O Box 627 Cornelius Oregon 97113 EMAIL tech Shellab com INTERNET http www Shellab com...

Страница 2: ...0 TROUBLESHOOTING SECTION 8 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS These units are general purpose air incubators for professional industrial or educational use where the preparation or testing...

Страница 3: ...e on the freight bill any damage found and enter your claim on the form supplied by the carrier 1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If necessary the...

Страница 4: ...own indicates that you should consult your manual for further description or discussion of a control or item 2 2 Indicates AC Power 2 3 Indicates Heating 2 4 Indicates Adjustable Temperature 2 5 Indic...

Страница 5: ...ifting Handling These units are heavy and care should be taken to use appropriate lifting devices sufficiently rated for these loads Units should only be lifted from their bottom surfaces Doors handle...

Страница 6: ...6 3 6 Shelf Placement Place the two shelves in the desired position See Figure 1 Figure 1...

Страница 7: ...oint temperatures 4 3 Heating Lamp This green pilot lamp is marked HEATING ACTIVATED It indicates when the element has been activated and the incubator is heating When set point is reached the pilot l...

Страница 8: ...ature range for this incubator is 5 C above ambient to 70 C The increments of the graduated dial should be used as reference points for example if 37 C is the desired set point start with an initial c...

Страница 9: ...ased bleaches or abrasives as they will damage the stainless steel interior Regular periodic cleaning is required Special care should be taken when cleaning around sensing heads to prevent damage WARN...

Страница 10: ...compare wiring to wiring schematic in manual especially around any areas recently worked on Unit will not heat over a temperature that is below set point 1 confirm that amperage and voltage match dat...

Страница 11: ...k fuse circuit breaker on unit or in wall 3 check all wiring connections esp around the on off switch Unit is smoking out of box This is not an uncommon occurrence when the protective coatings heat up...

Страница 12: ...A 3300515 Fuse Holder N A 3300501 Heating Element 2350562 2350500 HEATING Pilot Lamp 4650554 4650554 I O Switch 7850579 7850579 Inlet With Fuse Holder N A 4200505 Power Cord European N A 1800500 Power...

Страница 13: ...nsions WxDxH Weight Capacity Cubic Ft Temperature C Exterior Interior Shipping Net SMI1E EI1 SMI1E 2 EO1 2 16 75x17 75x22 25 12x12x14 53 lbs 44 lbs 1 0 Amb 5 70 SMI2E EI2 SMI2E 2 EI 2 21 5x18x25 5 17x...

Страница 14: ...14 WIRE DIAGRAM SMI1E E1 110V POWER SWITCH 7850579 HEATING LIGHT 4650554 120V 250W FUSE CONTROL THERMOSTAT 1750860 POWER CORD 1800529...

Страница 15: ...15 WIRE DIAGRAM SMI2E EI2 110V POWER SWITCH 7850579 HEATING LIGHT 4650554 120V 250W FUSE CONTROL THERMOSTAT 1750860 POWER CORD 1800529...

Страница 16: ...16 WIRE DIAGRAM SMI1E 2 EI1 2 220V 2800502 EMI FILTER 1 2 4 1 2 4 FUSED INLET 4200505 FUSE POWER SWITCH 7850579 TOP CONTROL 1750813 BOTTOM CONTROL 1750813 HEATING LIGHT 4650554 125 WATTS...

Страница 17: ...WIRE DIAGRAM SMI2E 2 EI2 2 220V 2800502 EMI FILTER 1 2 4 1 2 4 FUSED INLET 4200505 FUSE POWER SWITCH 7850579 TOP CONTROL 1750813 BOTTOM CONTROL 1750813 HEATING LIGHT 4650554 120V 125 WATTS 120V 125 WA...

Отзывы: