11
MECHANICAL
Door not sealing
1/ check physical condition of gasket
2/ assure that gasket is in original location
3/ Verify that door latch is operating correctly and not miss-aligned or
broken.
OTHER
Unit or wall fuse/circuit breaker is blown
1/ check wall power source.
2/ compare current draw and compare to specs on data plate.
3/ see what other loads are on the wall circuit.
4/ has the circuit breaker on the front panel been tripped? If so, see
section 4.4
Unit will not turn on
1/ check wall power source
2/ check fuse/circuit breaker on unit or in wall
3/ check all wiring connections, esp. around the on/off switch
Unit is smoking
– out of box
This is not an uncommon occurrence when the protective coatings heat
up during initial operation. Put unit under vent and run at full power for
one hour until the smoke dissipates.
Contamination in chamber
1/ see cleaning procedure in section 6.1
2/ develop and follow standard operating procedure for specific
application; include definition of cleaning technique and maintenance
schedule.
If following these troubleshooting suggestions does not solve the problem, call Customer Service for
assistance. Please see the manual cover for information on where to contact Customer Service.