33
|
P a g e
USER MAINTENANCE
Warning
: Disconnect the unit from its power supply prior to maintenance or cleaning of this unit.
Avertissement
: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le cordon
secteur de la source d'alimentation.
C
LEANING AND
D
ISINFECTING
•
Periodic cleaning is required.
•
Do not use spray-on cleaners or disinfectants. These can leak through openings and coat
electrical components.
•
Do not use cleaners or disinfectants that contain solvents capable of harming paint coatings
or stainless steel surfaces.
Do not use chlorine-based bleaches or abrasives; these will
damage the chamber liner.
•
Consult with the manufacturer or their agent if you have any doubts about the compatibility
of decontamination or cleaning agents with the parts of the equipment or with material
contained in it.
•
If a hazardous material or substance has spilled in the unit, immediately initiate your site
Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and
follow instructions per the site policy and procedures.
Warning
: Exercise caution if cleaning the unit with alcohol or flammable cleaners. Always allow the
unit to cool down to room temperature before cleaning and make sure all cleaning agents have
evaporated or otherwise been completely removed prior to putting the unit back into service.
Avertissement:
Soyez prudent lorsque vous nettoyez l'appareil avec de l'alcool ou des produits de
nettoyage inflammables. Laissez toujours refroidir l'appareil à la température ambiante avant le
nettoyage et assurez-vous que tous les produits de nettoyage se sont évaporés ou ont été
complètement enlevés avant de remettre l'appareil en service.
Cleaning
1.
Disconnect the unit from its power supply.
1.
Remove all removable accessory components such as shelving.
2.
Clean the unit with a mild soap and water solution, including all corners.
o
Do not
use an abrasive cleaner
. These will damage metal surfaces.
o
Do not use deionized water to rinse or clean with.
o
Take special care when cleaning around the temperature sensor probes in the
chamber to prevent damage. Do not clean the probes.
3.
Rinse with distilled water and wipe dry with a soft cloth.