16 |
P a g e
INSTALLATION
P
OWER
F
EED
W
IRING
The oven comes provided with an integral 6-inch (150 mm) wire braid consisting of:
•
14-gauge high-temperature (300°C) hot wire – Black
•
14-gauge high-temperature (300°C) hot wire – Red
•
14-gauge earth ground wire – Green/Yellow
The wires for power source connection should be Green/Yellow – Earth; Black – Hot; Red – Hot.
The oven must be earth grounded using the protective conductor terminal
(green with yellow
stripe) wire. Do not remove the protective conductor (earth connection). Removing the protective
conductor will negate the oven’s protections against potentially dangerous electric shocks and
create a potential fire hazard.
L
IFTING AND
H
ANDLING
The oven is heavy. Use appropriate lifting devices that are sufficiently rated for these loads. Follow
these guidelines when lifting the oven:
•
Lift the unit only from its bottom surface.
•
Doors, handles, and knobs are not adequate for lifting or stabilization.
•
Restrain the oven completely while lifting or transporting so it cannot tip.
•
Remove all moving parts, such as shelves and trays, and secure the door in the closed
position during transfers to prevent shifting and damage.