These units are built for use indoors at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
, at
no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating outside these conditions may
adversely affect the unit temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
•
Proximity to other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
R
EQUIRED
C
LEARANCES
These clearances are required to provide airflows for ventilation and cooling.
6 inches (152 mm)
of clearance is required on the sides and back of the unit.
12 inches (305 mm)
of headspace clearance is required if the oven exhaust is vented from the
workspace through a duct or other channeling.
24 inches (620 mm)
of headspace clearance is required between the exhaust vent and any
overhead cover or partition. This may be reduced to
12 inches (305 mm)
if ducting is
connected to the exhaust port.
Do not place objects on top of the oven. Exception: A power exhaust blower may be mounted
on the top exhaust vent.
6” (152 mm) from Fan
24” (620 mm)
12” (305 mm) headspace
with ducting attached
6” (152 mm)
6” (152 mm)
35” (889 mm)
12” (305 mm)