14
Controller on at all times - "locked-up"
1/ Adjust set point to room temperature. If the unit is still heating,
replace the solid state relay.
2/Turn unit off and on to reset.
3/ If cannot change any condition on the front panel, call Customer
Service.
Controller timer resets on its own
1/ Confirm that power from wall is consistent and within specifications.
2/ Call Customer Service with serial number.
Front panel displays are all off
1/ Check connections to the temperature display control board and
assure that all are tight and in the correct orientation.
2/ Check for wire damage.
Unit or wall fuse/circuit breaker is blown
1/ Check wall power source.
2/ Compare current draw and compare to specs on data plate.
3/ See what other loads are on the wall circuit.
Unit will not turn on
1/ Check wall power source.
2/ Check fuse/circuit breaker on unit or in wall.
3/ See if unit is on, e.g., fan or heater, and just controller is off.
4/ Check all wiring connections, especially around the on/off switch.
Unit is smoking
– Out of box
This is not an uncommon occurrence when first operating new units. Put
unit under vent and run at high temperature for one hour until smoke
disipates.
Contamination in chamber
1/ See cleaning procedure in section 7.0.
2/ Develop and follow standard operating procedure for specific
application; include definition of cleaning technique and maintenance
schedule.
Contamination in sample
1/ See “Contamination in chamber”.
2/ Reduce air flow in chamber by dampening down exhaust port; be
sure to verify adequate temperature uniformity at the reduced air flow.
3/ Protect open samples from areas of maximum air current, e.g., inlet
air ducts.