24 |
P a g e
OPERATION
C
ONDENSATION AND THE
D
EW
P
OINT
Condensation takes place whenever the humidity level in the incubator chamber reaches the dew
point. The dew point is the level of humidity at which the air cannot hold more water vapor. The
warmer the air, the more water vapor it can hold.
As the level of humidity rises in an incubation chamber, condensate will first appear on surfaces that
are cooler than the air temperature. Near the dew point, condensate forms on any item or exposed
surface even slightly cooler than the air. When the dew point is reached, condensate forms on nearly
all exposed surfaces.
Ambient relative humidity humidity exceeding 80% or overloading the incubation chamber with open
or breathable sample containers will likely result in condensation in the chamber, possible leaks
around the incubator. This may cause corrosion damage if allowed to continue for any significant
length of time.
Managing condensation primarily depends on either lowering the humidity level or increasing the air
temperature in the incubator chamber.
Note:
Rising or falling air pressure from the weather will adjust the dew point up and down in small
increments. If the relative humidity in the incubation chamber is already near the dew point,
barometric fluctuations may push it across the dew threshold.
Note:
Thin air at higher altitudes holds less humidity than the denser air found at or near sea level.
If excessive condensate has appeared in the incubation chamber, dry the chamber interior and
check the following.
•
Check the door gaskets for damage, wear, or signs of brittleness or dryness. Arrange for
replacement of the gaskets if damaged or excessively worn.
•
Verify the chamber access port is closed. The black, plastic shipping cap that came with the
unit should be installed on the outside of the incubator and not in the chamber.
•
Make sure samples on the shelves are evenly spaced to allow for good airflow.
•
Ensure the chamber door is closing and latching properly.
•
Are frequent or lengthy chamber door openings causing significant temperature disruptions
and chilling the chamber surfaces? If so, reduce the number of openings.
•
Are there are too many open or “breathable” containers of evaporating sample media in the
chamber? If so, reduce the number of open sample containers.
•
Does the ambient humidity in the room exceed the stated operating range of 80% relative
environmental humidity? If so, lower the room humidity.
•
Is the incubator exposed to an external flow of cold air such as an air-conditioning vent or a
door to a cooler hallway or adjacent room? Block or divert the air, or reposition the unit.
Содержание SMI11
Страница 1: ...SMI INCUBATORS 100 120 Voltage Installation Operation Manual SMI2 SMI6 SMI7 SMI11 SMI12...
Страница 2: ...2 P a g e Pictured on cover SMI6 SMI2 SMI7 SMI11 SMI12 2 Stacked SMI6s...
Страница 14: ...14 P a g e INSTALLATION...
Страница 16: ...16 P a g e GRAPHIC SYMBOLS...
Страница 18: ...18 P a g e CONTROL PANEL OVERVIEW...
Страница 32: ...32 P a g e USER MAINTENANCE...