OPERATION
S
ET THE
CO
2
S
ET
P
OINT
The incubator comes from the factory with the CO
2
set point set to Off. Set the set point percentage
to the gas concentration required by your study or production protocol.
LO - Low CO
2
When starting with no CO
2
in the chamber, the display will show “LO” until the incubator registers a
concentration greater than 0%. A few minutes of CO
2
injections may be required to build up a
sufficient gas concentration.
End of Procedure
1.
Navigate to the CO
2
Set Point Adjustment mode
Press and hold
either
Set CO
2
%
Set Point Adjustment Mode
Note: The display will automatically exit the adjustment mode after 5 seconds of
inactivity on the arrow keys, saving the last shown set point value.
2.
Set the CO
2
Set Point
Set CO
2
%
New Set Point
Adjust
Note: To turn the CO
2
system off, set the set point
to its lowest setting (OFF).
3.
Wait for 5 seconds after entering the Set Point
Set CO
2
%
The display will stop flashing. The set point is now saved in the controller.
The display will revert to showing the current chamber CO
2
concentration.
CO
2
Injecting
Set CO
2
%
Initial Set Point