OPERATION
Note:
Ensure that each incubation chamber in the SCO10A is connected to its own CO
2
supply.
S
TART A
F
LOW OF
CO
2
Gas Cylinder
Open the cylinder valve to supply CO
2
to the regulator.
Set the cylinder regulator flow setting.
•
Set the regulator to deliver 15 – 20 Pounds per Square Inch (psi) to the incubator
CO
2
in Port.
Never
exceed
20 psi
.
Building Supply System
Initiate a flow to supply 15 – 20 psi of CO
2
to the incubator through the Gas In port.
PSI
Megapascals
Kilopascals
Bar
15 – 20 psi
0.103 – 0.137 Mpa
103.42 – 137.89 Kpa
1.03 – 1.378 bar