USER MAINTENANCE (CONTINUED)
Warning
: Prior to maintenance or service on this unit, disconnect the power feed from the power
supply.
Avertissement
: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le
cordon secteur de la source d'alimentation.
C
LEANING AND
D
ISINFECTING
If a
hazardous material or substance
has spilled in the unit, immediately initiate your site
Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow
instructions per the site policy and procedures.
The BACTRON should be cleaned and disinfected prior to first use. Periodic cleaning and
disinfection are required to prevent microbiological contamination.
Do not use spray-on cleaners or disinfectants. These can leak through openings and coat electrical
components. Do not use cleaners or disinfectants that contain solvents capable of harming paint
coatings, acrylic glass, or stainless steel surfaces.
Do not use chlorine-based bleaches or
abrasives
—these will damage the chamber liner.
Consult with the manufacturer or their agent if
you have any doubts about the compatibility of decontamination or cleaning agents with the parts of
the equipment or with material contained in it.
Warning
: Never clean the unit with alcohol or flammable cleaners.
Avertissement:
Ne jamais nettoyer l'appareil à l'alcool ou avec des nettoyants inflammables.
Cleaning
Perform the steps below for a complete cleaning of the BACTRON interior:
1. Remove and clean the sleeve assemblies and all removable interior accessories, except for
the catalyst cartridges.
a.
Never clean the catalyst
using water, cleaning agents, or disinfectants
. See page
54 for how to clean the catalyst cartridge with heat and AMG.
b. Wash the arm port doors, sample dish racks, shelf spacers, pass box gaskets, and
sleeve assemblies with a mild soap and water solution.
2. Clean the workspace chamber, incubator, and pass box interiors with a mild soap and
water solution, including all corners.
a. Take special care when cleaning around and chamber power outlets to prevent
damage. Do not clean the pass box door alarm sensors (see Figure 33).
b.
Do not use chloride-based cleaners
except
Zephiran benzalkonium chloride solution.
Other types may have adverse effects on microbiological samples.
3. Rinse with distilled water and wipe dry with a soft cloth.
Do not use deionized water
.
See
paragraph in the Operation section.
4. Wipe down the interior surfaces with Zephiran. Allow the Zephiran to evaporate, do not
wipe up.
Содержание BACTRON300
Страница 2: ...2 P a g e This page left blank...
Страница 13: ...13 P a g e RECEIVING YOUR BACTRON CONTINUED This page left blank...
Страница 25: ...25 P a g e INSTALLATION CONTINUED This page left blank...
Страница 28: ...28 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank...
Страница 57: ...57 P a g e OPERATION CONTINUED This page left blank...
Страница 65: ...65 P a g e USER MAINTENANCE CONTINUED This page left blank...
Страница 85: ...85 P a g e ACCESSORIES CONTINUED...
Страница 86: ......