8.2 Using the AcuTrak Setup Menu
To enter the Setup Menu press and hold the key for 3 seconds. This displays
the
SEtuP
message on the main display.
Pressing the key takes the user to the
rAngE
function in the Setup menu
structure.
Note 1
: If the user has a setup menu password, then the user is prompted to enter
the correct password before continuing to the
rAngE
function
Note 2
: Most of the screens within the menu have an active 10 second time out.
Therefore, if no keys are pressed within this period the unit reverts automatically
to normal operation. In most cases where the 10 second time out occurs, changes
will not have been saved.
Table 1 Setup Menu contents
8.3 Sensor Range
The
rAngE
option allows the user to select the required sensor range. This option
is used to match the AcuTrak to the sensor connected to the unit. Press the key
while
rAngE
is displayed to enter the submenu. The currently selected range is
displayed. Select a different range by pressing the key to scroll through each
of the available ranges. Please see
Section 4.1.2 Available AcuDew Dewpoint
Measurement Ranges
.
ESC
Sensor Range
Contains a list of available ranges.
Configuration
Contains submenus for choosing engineering units and choice of
gas types.
Analogue output
Contains submenus to allow the user to fully configure the
Analogue output.
Alarms
Contains submenus to fully control behaviour of two
independent alarms.
Communications
Contains submenus for changing baud rate and address of the
instrument’s RS485 communication port.
Security
Contains submenus to set up passwords to control or limit
access certain features from unauthorized changes.
28