11 Technical aspects
11.3
Error messages
An error message is displayed in case of any failure during ope
-
ration. The error No. and error description will be indicated.
Error code LCD (error
description)
Cause
Remedy
F09
no adjustment:
no adjustment data
present
carry out an adjust-
ment
F10
adjustment
error:
zero point in the ppm
range
check test gas or re-
peat adjustment
F13
adjustment
error:
sensitivity in the ppm
range
check test gas or re-
peat adjustment
F51 – F54
system error
component error
witch instrument off
and on again or con-
tact SEWERIN service
F 59
voltage supply voltage outside the
permitted range
contact Sewerin serv-
ice only
F100
pump power
pump power too low
check filters in the in
-
strument and probes.
Please observe the
instructions in section
9.9: Adjusting pump
operation
Note:
If any other error codes appear, please contact
SEWERIN service!
Cause:
Remedy:
Programm defect
SEWERIN service
F09 – F15
temporary errors
F40, F100
clearable errors
F51 – 54, F59 – F62
system errors; switch instrument off