background image

25

Appendix I

User-Definable Units

USER-DEFINABLE UNITS

User-definable units are units of measure whose conversion factors are determined
by the user. Setra scales have two user-definable units of measure: X and T units.
These units may be used to weigh in dollars (given the weight of one dollar of
substance), length (given the weight of a known length of material), or volume (given
the weight of a known volume of material). Conversion factors can be entered
through the keyboard to be multiplied by the displayed weight, or divisors can be
entered to be divided into the displayed weight.

X UNITS

X units can be directly changed by entering a factor or divisor. To measure in X units,
proceed as follows:

1. Calculate the ratio you wish to enter. If your ratio is in grams per X unit, ounces

per X unit, pounds per X unit, etc., you will enter a divisor. Proceed with Steps 2
through 7.

If your ratio is in X units per gram, X units per ounce, X units per pound, etc.,
you will enter a factor. Proceed with Steps 2 through 7, skipping step 3.

2. Press the 

conv

 key until the display reads in the known units from which you

wish to convert.

3. If you are entering a divisor, press 

.

4. Enter the conversion ratio.

5. Press the 

conv

 key.

6. Place an object on the scale.

7. Wait for the “ok” symbol, then read the display. The reading given is in X units.

Содержание 140CP

Страница 1: ...ET TOTALIZING BY WEIGHT 14 BASIC COUNTING 15 COUNTING WITH THE TARE FUNCTION 15 COUNTING WITH THE GROSS FUNCTION 16 COUNTING WITH THE AVERAGE PIECE WEIGHT FUNCTION 16 COUNTING WITH ACCURACY ENHANCEMEN...

Страница 2: ...Scale Examples 34 Appendix V 36 Glossary of Terms 36 Appendix VI 37 RS 232 Serial Data Communications 37 Interfacing a Setra Scale to a Computer 37 Sending Commands to the Scale 37 Sending Alphanumer...

Страница 3: ...while using your scale Section Three Using Your Scale provides the actual instructions necessary to perform your scale s various functions Section Four Dual Scale Operation describes a special case w...

Страница 4: ...4 Figure 1 Components and features of the Setra platform scale...

Страница 5: ...and pressing in place until well seated The pan will be seated correctly if it is level on all sides 3 ConnectingyourscaletoanACoutlet To connect your scale to an AC outlet insert the appropriate end...

Страница 6: ...6 Figure 2 Detaching the controller...

Страница 7: ...r To detach the controller follow the procedure below refer to Figure 2 1 Remove the pan by grasping it with two hands on the sides and lifting 2 Loosen the controller attaching screws The controller...

Страница 8: ...cally runs through a display test showing all possible segments which might be displayed This test lasts approximately six seconds The scale next says HELLO and performs self diagnostics to ensure pro...

Страница 9: ...btracts the weight from the display and saves it in the tare register leaving the display to read net or contents weight gross Adds weight saved in tare register to net weight to display gross weight...

Страница 10: ...piece weight accum Reading shown is the amount in pieces or in weight stored in the accumulator register tare Reading shown is the weight stored in the tare register pc Reading shown is given in numbe...

Страница 11: ...The scale is recomputing the average piece weight to enhance accuracy dnOS Thescalehasbeenplacedintheself diagnosticmode Press setup cal the scale will reset itself to the factory defaults TArE CLEAr...

Страница 12: ...e register key you wish to recall e g press rcl tare CONVERTINGWEIGHINGUNITS Your scale is capable of weighing in pounds ounces grams X units T units carats pennyweights or troy ounces To convert from...

Страница 13: ...symbol then read the net weight from the display Note When you press the tare key the display ONLY is set to zero This is not the same as using the zero key When using the tare key the container weig...

Страница 14: ...both the gross and net weight is called net totalizing To net totalize weight use the following procedure 1 Press the key 2 Press the clear key 3 Press the zero key 4 Place a container on the pan 5 P...

Страница 15: ...this by taring out subtracting the weight of the container before it begins counting To tare out the weight of the container before counting pieces being placed in it use the following procedure 1 Pre...

Страница 16: ...again or pressing clear will return the display to net weight COUNTINGWITHTHEAVERAGE PIECE WEIGHTFUNCTION You can use your scale to count the number of pieces in a batch a handful of capacitors for ex...

Страница 17: ...ces on the scale 3 Enter the number of pieces you have placed on the scale 4 Press the count key Wait for the OK symbol then slowly add pieces until the displayed count is increased by 100 or less 5 P...

Страница 18: ...e following procedure 1 Press the then clear keys 2 Press the zero key and place a container on the pan 3 Press the tare key 4 Place a known number of pieces into the container 5 Enter the number of p...

Страница 19: ...Press the count key The scale will display the number of pieces equal to the minimum sample size required 7 Add pieces to continue counting 8 Wait for the ok symbol then read the display The scale no...

Страница 20: ...y The scale can be used to accumulate both weights and counts simultaneously Before accumulating press the clear keys to clear out all memory registers To accumulate a weight use the following procedu...

Страница 21: ...For example if 123456789 is entered the scale will display 456789 but will store and print 123456789 as explained in Printing Out Information The key may serve as a dash hyphen during I D number entry...

Страница 22: ...formation on a batch Note When using a printer other than the Setra 240P or 340P set the baud rate on your scale to match the printer see Appendix II It is possible to select the information to be pri...

Страница 23: ...items comprise a dual scale system a sampling scale a bulk counting scale and an interface cable The connectors and interface cable are available from Setra Prior to connecting the sampling scale and...

Страница 24: ...on the pan of the sampling scale 4 Enter that number into the sampling scale and press the count key 5 Enter the part number if any on the sampling scale and press the i d key 6 Press the apw key The...

Страница 25: ...into the displayed weight XUNITS X units can be directly changed by entering a factor or divisor To measure in X units proceed as follows 1 Calculate the ratio you wish to enter If your ratio is in gr...

Страница 26: ...is in T units per gram T units per ounce T units per pound etc you will enter a factor Proceed with Steps 2 through 10 skipping step 6 2 Press the conv key until the display reads in the units from wh...

Страница 27: ...de may be accessed by pressing the print key 4 Press the setup cal key the scale now asks if you wish to print the tare weight 5 Enter 1 if you wish to print the tare weight enter 0 if you do not 6 Pr...

Страница 28: ...ter 3333 2 Press the setup cal key The display will read CAL AC 3 Enter the desired accuracy in percent e g 99 9 4 Press the setup cal key Note For any given sample the scale automatically calculates...

Страница 29: ...isplay The scale asks if you wish to print the quantity count 14 Enter 0 and press the setup cal key Your scale is now setup to produce bar code labels with a Setra Auto Count 200 system For more info...

Страница 30: ...to another Refer to Section Four for complete instructions on the operation of a dual scale system Setting up a scale for dual scale operation disables the print function Pressing print will cause the...

Страница 31: ...r code label printer version 2 05 and lower No minimum sample size No minimum accuracy The beeper is enabled The INT mode is set to fast display update short integration time Span factory calibration...

Страница 32: ...below to change the functionality of the INT mode to either slow or fast display update 1 Press the key and enter 11110 This is the INT mode selection access code 2 Press the setup cal key The displa...

Страница 33: ...key 3 Press the zero key 4 Place the weight on the pan 5 Press the setup cal key SPANCALIBRATIONWITHASPECIFICWEIGHT To calibrate your scale with a specific known weight less than full scale use the f...

Страница 34: ...length of wire coiled on a spool Using the tare function you would first enter the tare weight of the empty spool Next using your user definable X units see Appendix I you would calculate the ratio of...

Страница 35: ...onuniform ball bearings using a small sample size You would enter the number of pieces on the pan followed by 0 The scale will display tenths of pieces Next you would add pieces If the digit to the ri...

Страница 36: ...ight of its container tare weight KITTING Repeatedly removing a given number of parts for inclusion in a kit e g removing the number of like transistors necessary to manufacture a radio NETWEIGHT The...

Страница 37: ...face hardware information can be found later in the appendix For information on connec tion and data type consult The RS 232 Interface Hardware section InterfacingaSetraScaletoaComputer The bidirectio...

Страница 38: ...ses v3 2 and higher this could cause a conflict The following setup allows the user to redefine E back to the remote scale com mand If E is redefined to the remote scale command exponential entry is n...

Страница 39: ...ollowing string to the scale string identifier S The ID START or introducer instructs the scale to ignore subsequent charac ters as commands The ID FINISH or terminator instructs the scale to interpre...

Страница 40: ...point depends on the precision of the number transmitted c1 is the mode character It describes what the number being transmitted represents The following is a list of all possible characters sent in...

Страница 41: ...ransmitted or the sequence rcl tare print was pressed the scale would send the contents of the tare register Using this mode a computer can read data from the scale only when specifically requested If...

Страница 42: ...n a given weight unit Using this function assures that numbers such as tare weight or average piece weight entered through the RS 232 interface are interpreted in the correct units For example if the...

Страница 43: ...toredinthismemory Asaresult thebaudrate and print format will not need to be reset each time the scale is powered on However if the continuous print mode is desired follow the procedure previously out...

Страница 44: ...mal buffer 15 characters Some computers requiring handshaking will need a connection between two pins on the computer s connector named DTR and DSR Data Terminal Ready and Data Set Ready CTS may also...

Страница 45: ...for label advance Designed for adhesive backed labels 3 1 2 x 15 16 and will include four lines of information per label 402645 INTERNAL DC BATTERY OPTION Must be ordered with scale Includes scale ci...

Страница 46: ...nd span calibration If these User Units and User Setups are not reentered the scale will default to factory settings If the scale displays ProtEC remove the stainless steel pan and flip the exposed me...

Страница 47: ...ices especially computers it may be necessary to cross the receive and transmit lines of the interface That is the receive line of the scale must connect to the transmit line of the external device Li...

Страница 48: ...d perform the following steps after removing power from the unit 1 Turn the scale over so the scale is resting on its pan 2 Remove the two Phillips head screws on the battery cover The battery cover i...

Страница 49: ...49 Figure 3 Replacing the optional battery pack...

Страница 50: ...le from the controller printer circuit board 5 Connect one end of the new cable to the controller board 6 Feed the new cable into the strain relief on the controller bottom cover 7 Replace the bottom...

Страница 51: ...51 Figure 4 Replacing the controller cable...

Страница 52: ...per second Response Time less than 2 seconds typical Counting Capacities 999 999 pieces Data Communications Bidirectional RS 232 Selectable Baud Rates 300 2400 600 4800 1200 9600 Display 6 digit LCD...

Страница 53: ...lace scales which have been repaired by unauthorized parties abused improperly installed altered or otherwise misused or damaged even if by accident in any way Setra will not be responsible for any di...

Страница 54: ...h shielded cables This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of...

Страница 55: ...40 41 42 43 44 47 Continuous print 40 41 42 43 Conv 9 12 25 33 34 38 Count 9 11 15 16 17 19 20 21 23 24 27 29 32 34 35 38 45 Dual scale 3 23 24 30 46 Examples 17 Factory defaults 31 Firmware 30 Flash...

Страница 56: ...le 17 Sampling 23 47 Setup cal 9 38 Specifications 52 Storing information 39 String format 41 T units 10 12 25 26 40 Tare 13 34 36 38 40 Tone 31 43 46 UnAbLE 8 11 46 UPdAtE 17 Update 17 Verify functio...

Отзывы: