11
DISPLAY MESSAGES
bUSY
The scale is developing a stable reading.
UnAbLE
The scale is unable to complete your requested function. Press the
clear
key and select another function.
OFLO
The counting display has overloaded beyond 999,999 pieces. Re-
move some of the pieces and count in smaller lots.
FLASHING
The weight on the pan exceeds the capacity of the scale. Remove
DISPLAY
some of the weight from the pan. If you are counting, count in smaller
lots.
UPdAtE
The scale is recomputing the average piece weight to enhance
accuracy.
dnOS
The scale has been placed in the self-diagnostic mode. Press
setup/cal
,
the scale will reset itself to the factory defaults.
TArE CLEAr
If a value is stored in the tare register when the
zero
key is pressed, this
message will appear warning that the tare value has been cleared.
Add nnn Pc
The scale is requesting an additional number of pieces (nnn) to satisfy
minimum accuracy and/or minimum sample size setup requirements.
Sub nnn Pc
When kitting, the scale is requesting an additional number of pieces
(nnn) be removed to satisfy minimum accuracy and/or minimum
sample size setup requirements.