![Servomex SERVOTOUGH Oxy Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065043.webp)
01910001B_8
© Servomex Group Limited. 2020
31
1.
Refer t
Use the 2 mm hexagonal key supplied to loosen the locking screw
which secures the cover.
2.
Unscrew
the cover; that is, turn it an�clockwise. If necessary, fit a suitable metal bar
between the protruding fins on the front of the cover to provide addi�onal leverage.
3.
When
the cover is fully disengaged from the body of the analyser, remove it and place it
carefully on a clean surface: see the cau�on above.
5.3
Electrical supply and interface signal connections
5.3.1
Electrical
safety
Ensure that the electrical installa�on of the analyser
conforms with all applicable
local and na�onal hazardous area and electrical safety requirements.
Obey the safety instruc�ons given below when you install the analyser; if you do not,
the
analyser
cer�fica�on may be invalidated, the analyser
may not operate correctly,
or it may be damaged.
•
All cables connected to the power/signal compartment must be connected to
equipment which is not supplied from, or contain in normal or abnormal opera�ng
condi�ons,
a source of poten�al with respect to earth (ground) higher than 253 Vac or
253 Vdc. As such it is recommended that the analyser
is connected to a "grounded
neutral" supply.
•
The
analyser
does not incorporate an integral on/off switch. You must provide a means
of externally isola�ng the electrical supply from the analyser: use a suitable switch or
circuit breaker located close to the analyser, clearly marked as the disconnec�ng device
for the analyser. This switch/circuit breaker must be easily operated with no
obstruc�ons.
•
The electrical supply circuit must incorporate a suitable fuse or over
-
current protec�on
device, set to or rated at no more
than 10 A.
•
The
analyser
must be connected to an external protec�ve earthing system, using the
Protec�ve Earth on terminal P101.
•
The
analys
er
s Earth (ground) connec�on located on the base of the analyser
must be
equipoten�ally bonded locally to the analyser
.
•
Ensure that your electrical supply can provide the necessary maximum power
consump�on: refer to Sec�o
•
The volt
-
free relay contacts are isolated from the analyser
mains circuits and from each
other, only 30 Vdc 1A max signal voltage circuits can be connected to the contacts.
•
The
mA output and RS485 terminals are separated from the analyser
mains circuits by
reinforced insula�on. The terminals must only be connected to circuits that are
themselves separated from mains voltages by at least reinforced insula�on.
•
All signal and electrical supply cables must be rated for temperatures of 70
°C
(158
°F
) or
h
igher.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...