SERVOTOUGH LaserExact Monitor, Operator Manual 02950001A_1
ELECTRICAL
CONNECTIONS
61
6.9.2
Inverted configuration of relay-3, «Gas alarm»
This is defined by setting the parameter “Gas conc alarm level” < 0 (negative
value) for gas-1. Inverted relay-3 function then also applies to gas-2 and gas-3
alarm levels (if applicable) irrespective of their sign. Alarm is signified when the
measured gas concentration > abs(Gas conc alarm level) for one or more of the
measured gases. Inverted configuration:
Alarm ON = Relay-3 conducting (X)
Alarm OFF = Relay-3 isolating (O).
6.9.3
Maintenance configuration of relay-3
Maintenance indication is done by relay no. 3 by setting the parameter “Gas conc
alarm level” = -10
20
(= -1E20)
for all gaslines, N
, of the instrument (see Section
4.3.6). This ensures that «Gas Alarm» is disabled for all realistic gas
concentrations and does not interfere with the maintenance function.
“Normal” measuring mode: Relay-3 isolating (O)
“Zero”, “Span”, or “Maintenance” mode: Relay-3 conducting (X).
6.9.4
Inverting relay-3 output by hardware switch
The AUX-board switch SW1-3/C is by default set to position 3 (REL3_NC). This
gives the relay outputs as described in the sections above. When the AUX-board
switch SW1-3/C is set to position C (REL3_NO), the relay function is interchanged.
This also applies to operating mode “Power off” (see also Section 6.3).
Содержание SERVOTOUGH LaserExact 2950
Страница 2: ...This page intentionally blank...
Страница 6: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 IV This page intentionally blank...
Страница 69: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 63...
Страница 70: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 64 APPENDIX...
Страница 71: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 65...