Plasma Gas Analyzer
49
02001001A_9
5.8.7 Averaging Number
Figure 40 - AVERAGING NUMBER
Minimum: 1
Maximum: 25
Resolution: 1
The software may execute a moving average on the calculated ppm value. This will also affect
the response time of the system. The software uses this parameter to determine the number of
sample PPM values that are used in its moving average sub-routine. The number of samples
ranges from 1 to 25. When entering 1, there is no moving average.
For example, in a 5 point moving average filter, the displayed actual process valve will be:
Ppm = p (1) + p (2) + p (3) + p (4) + p (5)
5
Where p (5) is the new measure done and p (1) to p (4) are the four previous ones.
<<<AVERAGING NUMBER>>>
Averaging:3
Actual:1
F3:NEXT F4:RET
Содержание SERVOPRO Plasma
Страница 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Страница 2: ...This page intentionally blank...
Страница 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Страница 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Страница 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Страница 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Страница 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Страница 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Страница 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...